Remote Access To Webmin Using A Service Pc Over The Ix Network - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

The Login to Webmin dialog box displays.
9.
Continue to
Closing Webmin
1.
You can exit the monitor's maintenance mode any time by pressing any keypad or
keyboard key, using the Trim Knob, using the touchscreen or clicking a mouse button. The
monitor will automatically restore the original network configuration to the S/5 or MC
Network and return to the normal monitoring mode.
2.
Restore the original network settings in the service PC.
3.
Disconnect the crossover cable from the monitor and from the service PC.
4.
Reconnect the monitor back to the S/5 or MC Network if applicable.
4.4 Remote access to Webmin using a service PC over the
IX Network
NOTE: This connection method is supported only if the monitor is equipped with the Advanced
Interface Board that provides both the IX and the MC connectors and the monitor is connected
to a live IX Network.
1.
Connect a service PC to the IX Network using a standard network cable.
2.
In patient monitor, select Monitor Setup > Service. The local browser opens and displays
the Login to Webmin dialog box.
3.
Record the IX IP address of the patient monitor:
IX IP address: ______________________
4.5. Login to
Webmin.
Using Webmin service interface
2081903-001