12.3 Main unit disassembly

12.3.1 Replacing the mains fuses
184
2081903-001
The mains fuses are situated on the back of the
monitor, below the mains connector.
1. Lever the screwdriver against the cable holder
to release the fuse holder and pull it out of the
frame.
2. Pull the fuses out of the fuse holder and insert
new ones.
3. Push the fuse holder all the way in until it
snaps to its place.