Reassemble in reverse order.
-
Attach the 4 screws that mount the AC/DC board [Torque 1.6 Nm].
-
Ensure the cable clamp and the spring are in place.
-
Reattach the housing back cover so that the 4 snaps locks.
-
Attach the 1 screw that mounts the housing back cover.
-
Reconnect the AC/DC cable and the protective earth cable to the AC/DC board.
-
Place the DC/DC cable in the cable holder.

12.3.11 Detaching the Base unit

Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
d.
12.3.5. Disassembling the main unit into Rear and Front units
e.
12.4.1. Detaching the Front Unit Assembly
f.
12.4.2. Detaching the LCD Display Unit
Disassembly and reassembly
6. Remove the four T10 screws mounting the
AC/DC board to the back plate.
[Torque 1.6 Nm]
Hold the AC/DC board to prevent it from
falling.