Functional check
IX printers
Insite with EXC
Recorder
Synchronization connector test
Test completion
Notes
Used service parts
Signature
Observed result
Acceptance criteria
Test page is printed to the selected
printer.
The patient monitor is active on Insite
with EXC back office.
The header line contains the correct
information. The grid is clear. The
waveforms labels appear in the
printout as configured.
ECG, Arterial BP, and Marker Out
signals are correct.
Maintenance check form
PASS
N.A.
FAIL
Date
B - 3 (4)
2081903-001