Canceling Pending Host Software Activation - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
activation is completed and the parameter module has restarted. This may take up to 15
minutes. Do not shut down the patient monitor or disconnect the parameter modules.
The parameter module restarts automatically after the software activation is complete.
4.
After the parameter module restarts, verify that the software activation is successful and
the parameter module runs the activated software.

7.18.3 Canceling pending host software activation

You can cancel a pending host software activation at any time while the message Software
activation after next discharge/after next case end is shown on the display.
1.
Log in to Webmin.
2.
Select Configuration > Software Management.
The Software Management window opens.
3.
Select Host Software from the Software List and select
4.
In the Host Software window, below Status, verify that the Software activation pending
is Yes.
5.
Select Cancel.
The pending software activation is cancelled and the patient monitor continues running the
current software version.
7.18.4 Erasing an inactive software version.
You can erase an inactive host software version from the patient monitor. Erasing can prevent
the activation of a wrong software version by mistake.
1.
Log in to Webmin.
2.
Select Configuration > Software Management.
The Software Management window opens.
3.
Select the software that you want to erase and select Next.
The selected software activation window appears.
4.
Select Erase to erase the inactive software.