12.3.6 Detaching the uDOM

Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
d.
12.3.5. Disassembling the main unit into Rear and Front units
uDOM replacement:
Contact GE Service if you doubt that the uDOM is defective and needs to be replaced. Some of the
configuration steps required after uDOM replacement can be performed only by GE service.
Provide the following information:
a) Monitor Type
b) Serial Number
c) Software version installed, if available.
190
2081903-001
1. Detach the uDOM from the CPU connector.
NOTE: uDOM, disk on module is the permanent
memory of the patient monitor. The software,
licenses and clinical and service configurations
are stored to the uDOM.
Reassemble in reverse order.