CARESCAPE Monitor B650
When reassembling, the recorder unit can be installed into the Module Frame as a whole.
NOTE: Ensure that the recorder unit is firmly installed to the recorder connector and both snaps in the EMC
cover lock properly.
12.2.3 Detaching the PDM docking mechanism
Disassemble first:
a.
12.2.1. Detaching the module frame front cover
•
Reassemble in reverse order.
180
2081903-001
1. Release the two latches on the inner left side
of the Module Frame by first pressing the
snaps and then sliding the latches towards
you.
2. Turn the PDM Docking Mechanism to a 30°
angle to detach it from the Module frame flex
board connector.