Detaching The Pdm Docking Mechanism - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
When reassembling, the recorder unit can be installed into the Module Frame as a whole.
NOTE: Ensure that the recorder unit is firmly installed to the recorder connector and both snaps in the EMC
cover lock properly.

12.2.3 Detaching the PDM docking mechanism

Disassemble first:
a.
12.2.1. Detaching the module frame front cover
Reassemble in reverse order.
180
2081903-001
1. Release the two latches on the inner left side
of the Module Frame by first pressing the
snaps and then sliding the latches towards
you.
2. Turn the PDM Docking Mechanism to a 30°
angle to detach it from the Module frame flex
board connector.