GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 89

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

a.
In the Position column, type the beginning position of the field in the data string
(from 1 to 300).
b.
In the Length column, type the number of characters (from 1 to 99) that the field
contains.
2.
For Gender Format, select Fixed or Configured.
If you select Configured:
a.
Type the character that identifies Male.
b.
Type the character that identifies Female.
3.
Below Fixed Option, select the applicable value:
Item
Item selection in the top part of
the screen
Height Unit
Both Height and Height Unit
Height only
Weight Unit
Both Weight and Weight Unit
Weight only
Age Unit
Both Age and Age Unit
Age only
4.
Scroll to the bottom of the window, and select Save.
All changes take effect immediately.
Example of Length Delimited Parser information
In this example, the barcode contains 10 items. The following table lists the starting position
and length of each item.
Item
MRN
First Name
Last Name
Day of Birth
Month of Birth
Year of Birth
Age
Age Unit
Gender
Height
The following sample shows the corresponding entries in the Admit/Discharge Configuration
window.
Starting Position
1
11
21
46
48
50
36
41
54
55
Fixed Option selection
Non-Fixed.
Select value from drop down list.
Non-Fixed.
Select value from drop down list.
Non-Fixed.
Select value from drop down list.
Length
10
10
15
2
2
4
5
5
1
5
Configuration