NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
Checkout procedure for wireless patient monitors
Check each wireless patient monitor according to the following procedure.
The purpose of this test is to ensure that each wireless patient monitor is correctly configured
1.
Disconnect the network cable from the MC Network connector, if connected.
2.
Check that the wireless network symbol and the signal strength symbol are shown
and that the signal strength is adequate.
3.
Select Data & Pages – Other Patients and select a patient from the list.
4.
Select View and verify that a window with parameters from another patient displays on
the left side of the screen. Check that the waveforms are continuous and there is no data
loss.
5.
Select Close View to close the window.
6.
Log in to Webmin.
7.
Select Diagnostics > Ping.
8.
Type the IP address of the patient monitor you just viewed and select ping. Check that the
patient monitor replies to the ping command, there are no lost packets and the maximum
latency is 250 ms.
NOTE: The patient monitor withstands a maximum packet loss of 5 packets per 1 million
and maximum latency of 250 ms without performance degradation.
9.
Select
10. Reconnect the MC Network cable back to the MC Network connector, if applicable.
Verification procedure for wireless MC Network infrastructure
Verify the wireless MC Network infrastructure according to the procedure described in
Appendix C.
The purpose of this test is to ensure the operation of the wireless network infrastructure
coverage area with a wireless CARESCAPE monitor.

8.3.11 IX printers

Perform the following test only if the patient monitor is connected to a printer in the IX Network
and you did not print a test page while you configured the IX printer.
1.
Log in to Webmin.
2.
Select Configuration > Printers > Print Test Page.
3.
Select the IX Printer to test.
4.
Select Save.
5.
Verify that the test page was printed to the selected printer.
6.
Repeat steps 3 to 5 for all connected IX printers.