Basic Service Manual
Part Number: 5194296-100
Revision: 10

Advertisement

Table of Contents
loading

Summary of Contents for GE HEALTHCARE Logiq Book XP series

  • Page 1 GE Healthcare LOGIQ BOOK XP Series Basic Service Manual Part Number: 5194296-100 Revision: 10...
  • Page 3 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Important Precautions THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY. • IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES.
  • Page 4 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
  • Page 5 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES. • KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST. HOIATUS • ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA (ET) TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
  • Page 6 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN SKILIÐ...
  • Page 7 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS IZLASĪŠANAS UN SAPRAŠANAS.
  • Page 8 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
  • Page 9 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA. •...
  • Page 10 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
  • Page 11 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL (ZH-CN) (KO)
  • Page 12: Part Number

    Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers.
  • Page 13 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 14 GE H EALTHCARE 5194296-100, R LOGIQ BOOK XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Revision History Revision Date Reason for change 2007/06/25 Initial Release. 2007/12/19 Add one Caution for CMOS battery 2008/09/18 Add one label for battery 2008/12/04 Delete section “Using a Phantom” 2010/09/15 Update the section “Important Precaution”...
  • Page 15: Table Of Contents

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 16 Electrical Requirements ........2 - 2 LOGIQ Book XP Series Power Requirements ....2 - 2 Inrush Current .
  • Page 17 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL CHAPTER 3 Installation Overview............3 - 1 Purpose of Chapter 3 .
  • Page 18 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Loading the System Software ......... 3 - 27 Loading the System Software with CD .
  • Page 19 Control Panel ......... 4 - 7 LOGIQ Book XP Series SoftMenu Key Tour ....4 - 8 Monitor Display .
  • Page 20 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Report* Pages ......... . 4 - 23 Probe/Connectors Usage .
  • Page 21 Service Home Page ......... . 5 - 9 Error Logs Tab (Not available on LOGIQ Book XP Series now) ..5 - 10 Logs .
  • Page 22 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of this chapter 6 .
  • Page 23 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
  • Page 24 ANUAL LOGIQ Book XP Series Diagnostic Descriptions......7 - 19 Troubleshooting ..........7 - 19 FRU Test .
  • Page 25 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview............8 - 1 Purpose of Chapter 8 .
  • Page 26 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL CHAPTER 9 Renewal Parts Overview ............9 - 1 Purpose of Chapter 9 .
  • Page 27 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview............10 - 1 Periodic Maintenance Inspections .
  • Page 28 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Probe Leakage Current Test ........10 - 16 Definition .
  • Page 29: Overview

    Customer Assistance 1-1-3 Purpose of Service Manual This Service Manual provides service information for the LOGIQ Book XP Series Ultrasound Scanning System. It contains the following chapters: 1.) Chapter 1 - Introduction: Contains a content summary and warnings. 2.) Chapter 2 - Pre Installation: Contains pre-installation requirements for the LOGIQ Book XP Series.
  • Page 30: Typical Users Of The Basic Service Manual

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-1-4 Typical Users of the Basic Service Manual • Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ Book XP...
  • Page 31: Important Conventions

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
  • Page 32: Standard Hazard Icons

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 33: Product Icons

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-4 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION...
  • Page 34 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 1-4 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “Consult accompanying documents” is Various intended to alert the user to refer to the operator manual or other instructions when complete information cannot be provided on the label.
  • Page 35 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 1-4 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION Input Rating Plate For use with adapter model TWADP Rating Plate Description Rating Plate Type CF Defib-Proof Applied Part (heart in the box with paddle) symbol is in ECG Module accordance with IEC 60878-02-06.
  • Page 36 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 1-4 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION The separate collection symbol is affixed to a battery, or its packaging, to advise you that the battery must be recycled or disposed of in accordance with local or country laws.
  • Page 37: Safety Considerations

    1-3-2 Human Safety Operating personnel must not remove the system covers. Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ Book XP Series Training are authorized to service the equipment. 1-3-3 Mechanical Safety...
  • Page 38: Electrical Safety

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-3-4 Electrical Safety To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with protective ground.
  • Page 39: Battery Safety

    ESD damage. • In the case of longer non-use of the LOGIQ Book XP Series, please make sure the battery is removed. To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,...
  • Page 40: Dangerous Procedure Warnings

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
  • Page 41: Lockout/Tagout Requirements (For Usa Only)

    6.) Control all stored and residual energy. 7.) Verify isolation. All potentially hazardous stored or residual energy is relieved. Energy Control and Power Lockout for LOGIQ Book XP Series NOTICE WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS: 1.
  • Page 42: Returning/Shipping Probes And Repair Parts

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-3-9 Returning/Shipping Probes and Repair Parts Equipment being returned must be clean and free of blood and other infectious substances. GEMS policy states that body fluids must be properly removed from any part or equipment prior to shipment.
  • Page 43: Emc, Emi, And Esd

    1-4-2 CE Compliance The LOGIQ Book XP Series unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
  • Page 44: Customer Assistance

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 1-5 Customer Assistance 1-5-1 Contact Information If this equipment does not work as indicated in this service manual or in the User Manual, or if you require additional assistance, please contact the local distributor or appropriate support resource, as listed below.
  • Page 45: System Manufacturer

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 1-5-2 System Manufacturer Table 1-6 System Manufacturer Manufacturer Phone Number GE Medical Systems (China) Co., Ltd. TEL: +86 510-85225888 No.19, Changjiang Road, Wuxi National Hi-Tech Dev. Zone, Jiangsu, P.R. FAX: +86 510-85226688 China 214028 GE Ultrasound Korea:...
  • Page 46 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL This page was intentionally left blank. 1-18 Section 1-5 - Customer Assistance...
  • Page 47: Pre Installation

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Chapter 2 Pre Installation Section 2-1 Overview 2-1-1 Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ Book XP Series.
  • Page 48: General Console Requirements

    ERVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-2 Environmental Requirements for LOGIQ Book XP Series Systems Operational Storage Transport Temperature 10 - 40 degree C -5 - 50 degree C -5 - 50 degree C...
  • Page 49: Inrush Current

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 2-2-2-2 Inrush Current Inrush current is not a factor to consider due to the inrush current limiting properties of the power supplies. Table 2-4 Inrush Current Inrush Current Voltage Console Only...
  • Page 50: Emi Limitations

    Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ Book XP Series complies with limits as stated on the EMC label. However there is no guarantee that interface will not occur in a particular installation.
  • Page 51: Scan Probe Environmental Requirements

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 40° C Storage:-10° to 50° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32). SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF NOTICE -10 TO + 50 degrees C.
  • Page 52: Facility Needs

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 2-3 Facility Needs 2-3-1 Recommended Ultrasound Room Layout 2-3-1-1 Purchaser Responsibilities The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
  • Page 53: Required Features

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system.
  • Page 54: Recommended And Alternate Ultrasound Room Layout

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 2-3-3-1 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: 18 IN . LINEN SUPPLY (46 C M ) D E D I C AT E D P O W E R R E C EP TA C LE...
  • Page 55: Networking Pre-Installation Requirements

    Information must include: • A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ Book XP Series. • The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
  • Page 56 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 2-3-4-4 DICOM* Option Pre-installation Requirements (cont’d) LOGIQ Book IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2...
  • Page 57: Overview

    * This function is option on LOGIQ Book XP Series. NOTICE The LOGIQ Book XP Series installation and functional checkout will take approximately one hour. LOGIQ Book XP Series consoles with optional equipment may take slightly longer. Chapter 3 Installation...
  • Page 58: Installation Warnings

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-1-3 Installation Warnings 1.) There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service personnel should carry out servicing.
  • Page 59: Safety Reminders

    Series and kept near the unit for quick reference. ACOUSTIC OUTPUT HAZARD DANGER Although the ultrasound energy transmitted from the LOGIQ Book XP Series probe is within FDA limits, avoid unnecessary exposure. Ultrasound energy can produce heat and mechanical damage...
  • Page 60: Receiving And Unpacking The Equipment

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
  • Page 61 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 3.) Take out the Paper pad. 4.) Take out console together with 2 interleavers. 5.) Take out the interleavers beside Accessories Package. 6.) Take out Accessories Package..
  • Page 62 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment (cont’d) 7.) Remove 2 interleavers. 8.) Remove plastic bag. Interleaver Plastic bag Interleaver Console Figure 3-3 Removing interleavers and plastic bag Section 3-2 - Receiving and Unpacking the Equipment...
  • Page 63: Moving Into Position

    Do not lift the console by holding any covers. In general, a single adult can move the LOGIQ Book XP Series. Before moving, store all loose parts in original accessory box or in back pack. Return probes to original box.
  • Page 64: Preparing For Installation

    Verify that the system is set to the correct voltage. The Voltage settings for the LOGIQ Book XP Series System is found on a label located on the AC adapter. • 220-240VAC(China); 100-120VAC(USA/Japan); 220-240VAC(Europe, Latin America) Connecting a LOGIQ Book XP Series System to the wrong voltage level will most likely WARNING WARNING destroy the system.
  • Page 65: Completing The Installation

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 66: Turn On The System

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-5-1-2 Turn on the system Press the Power On/Off switch at the front of the system once. Power On/Off switch Figure 3-6 Power On/Off Switch When the Power On/Off switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the system.
  • Page 67: System Shutdown

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-5-2-2 System Shutdown Disconnect the Mains Power Cable if necessary. For example: Relocating the system. 3-5-3 Transducer Connection 1.) Carefully open the system LCD display, plug the probe connector into the probe port, then lock the probe latch upward.
  • Page 68: System Configuration

    ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the LOGIQ Book XP Series console are summarized in Figure 3-9 on page 3-12 Table 3-4 Physical Dimensions of LOGIQ Book XP Series Height Width...
  • Page 69: Electrical Specifications

    EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-6-2 Electrical Specifications Table 3-5 Electrical Specifications for LOGIQ Book XP Series Adapter Voltage Tolerances Current Frequency ADM-9020M-GE 100-240 VAC +/-10% 1.08-0.45 A 50/60Hz 3-6-3 Approved peripherals...
  • Page 70: Connecting Cables

    5135661 For USB Printer & USB CD-RW 3-6-5 Peripherals/Accessories Connector Panel LOGIQ Book XP Series peripherals and accessories can be properly connected using the rear connector panel. 3-6-5-1 Rear Panel Connector Located on the rear panel are video input and output connectors, audio input and output, camera expose connectors, footswitch connector, power connector, and control connections for printer, and service tools.
  • Page 71 3.) Two interchangeable USB Port (Digital Printer, CD-RW and/or Footswitch, etc.) 4.) Port for DC input (AC Adapter) The USB devices should be connected to LOGIQ Book XP Series first, power on USB devices before NOTICE turning LOGIQ Book XP Series to work.
  • Page 72 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-6-5-1 Rear Panel Connector (cont’d) 1. Pin Assignment of DC/DC input Connector: 3 Pin, Female Table 3-8 Pin Assignments of DC/DC input Pin No. Signal Pin No.
  • Page 73 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-6-5-1 Rear Panel Connector (cont’d) 4. Pin Assignment of PCMCIA Table 3-12 Pin Assignments of PCMCIA Pin No. Signal Pin No. Signal DGND DGND BCDATA3 BCD1# BCDATA4 BCDATA11...
  • Page 74: Connect Peripherals

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 3-12 Pin Assignments of PCMCIA Pin No. Signal Pin No. Signal BCADR0 BBCD1 BCDATA0 BCDATA8 BCDATA1 BCDATA9 BCDATA2 BCDATA10 BIOCS16# BCD2# DGND DGND 3-6-5-2 Connect peripherals A.) Connect B/W printer to the system.
  • Page 75: Connect Peripherals

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-6-5-2 Connect peripherals (cont’d) B.) Connect color printer to the system. Color Printer can be properly connected using USB Port1 or USB Port 2(Figure 3-12 on page 3-19 Figure 3-12 Connect digital color printer to the system HP470 Deskjet Color Printer and HP K5400 Officejet Color Printer must connect with PIT (Printer NOTICE...
  • Page 76 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-6-5-2 Connect peripherals (cont’d) C.) Connect CD-RW to the system. CD-RW can be properly connected using USB Port1 or USB Port2. Figure 3-13 Connect CD-RW to the system Iomega CD-RW drive recommended media list Media issues are common throughout the CD-RW drive industry.
  • Page 77 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 3-13 Iomega CD-RW drive recommended media list Brand Name Type Product Code Speed Kodak CDR Gold Ultima 16x-1x Mitsui CDR74 32x-1x Mitsui CDR80 32x-1x Mitsubishi Chemical Ultra CDRW Mitsubishi Chemical...
  • Page 78 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-6-5-2 Connect peripherals (cont’d) D.) Connect Footswitch to the system. Footswitch can be properly connected using USB Port1 or USB Port2. Footswitch FSU 1000 Footswitch FSU 2001 Figure 3-14 Connect Footswitch to the system E.) Connect Wireless LAN Card to the system.
  • Page 79 Please refer to the operation manual of each peripheral for information needed by the user to operate the system safely. For detailed installation information, please refer to the LOGIQ Book XP Series Peripheral Installation Instruction manual 5194298-100. Chapter 3 Installation...
  • Page 80: Available Probes

    ASIC ERVICE ANUAL 3-6-6 Available Probes See in specification in the LOGIQ Book XP Series User Reference Manual for Probes and intended use. Table 3-14 List of Probes on LOGIQ Book XP/XP PRO Probe Name Material of Headshell Area of Using...
  • Page 81: Software/Option Configuration

    PERIPHERAL VASCULAR Section 3-7 Software/Option Configuration Refer to the LOGIQ Book XP Series Basic User Manual, Chapter 16, Customizing Your System for information on configuring items like Hospital, Department, Language, Units (of measure), Date, Time and Date Format. For information on configuring Software Options, Refer to the LOGIQ Book XP Series Basic User Manual, Chapter 16, Customizing Your System.
  • Page 82: Connectivity Installation Worksheet

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-8 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: REV: LOGIQ SN: Type: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title: IP Settings Remote Archive Setup...
  • Page 83: Loading The System Software

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-9 Loading the System Software NOTE: System Software may also be referred to as the Base Image or Ghost. While the software install procedure is designed to preserved data, you should save WARNING WARNING any patient data, images, system setups and customer presets to CD, DVD, USB Flash...
  • Page 84: Loading The System Software With Cd

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-1 Loading the System Software with CD 3-9-1-1 Loading Base Image Software (For R2.2.1 or lower) NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers.
  • Page 85 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-1-2 Upgrading Application Software (For R2.2.1 or lower) (cont’d) 5.) Press any key on the Keyboard to continue the steps. 6.) Press any key to continue when below message display as shown in Figure 3-19 on page 3- Figure 3-20 Update message 7.) System CD will be loaded as shown in...
  • Page 86: Upgrading Application Software (For R2.2.1 Or Lower)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-1-2 Upgrading Application Software (For R2.2.1 or lower) While the software install procedure is designed to preserve data, you should save any WARNING WARNING patient data, images, system setups to a CD-RW or hardcopy. 1.) Disconnect the system from the network.
  • Page 87 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-1-2 Upgrading Application Software (For R2.2.1 or lower) (cont’d) Figure 3-25 Software Load Complete 10.)When it completes loading, the system will reboot. 11.)Remove the CD from the drive. If possible, while the system is rebooting remove the CD from the drive.
  • Page 88 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-1-2 Upgrading Application Software (For R2.2.1 or lower) (cont’d) 15.)Click on the SHUT DOWN button to shut down the system after FPGA upgrade completing. Figure 3-28 FPGA Upgrade 3-32 Section 3-9 - Loading the System Software...
  • Page 89: Loading The System Software With Dvd (For R2.2.2)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-2 Loading the system software with DVD (For R2.2.2) While the software install procedure is designed to preserve data, you should save any WARNING WARNING patient data, images, system setups to CD, DVD, USB Flash Drive, or USB Hard Disk.
  • Page 90 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-2 Loading the system software with DVD (For R2.2.2) (cont’d) 7.) Press any key on the keyboard to continue the steps when message displays as shown in Figure 3-31 on page 3-34 Figure 3-31 Software upgrade continuing 3-34...
  • Page 91 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-2 Loading the system software with DVD (For R2.2.2) (cont’d) 8.) The system provides three options [a], [b] and [c] as shown in Figure 3-32 on page 3-35 To select [a], the complete disk will be loaded.
  • Page 92 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-2 Loading the system software with DVD (For R2.2.2) (cont’d) 10.)When the message displays as shown in Figure 3-33 on page 3-36 , input “Yes“ and press on “Enter“...
  • Page 93 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-2 Loading the system software with DVD (For R2.2.2) (cont’d) 12.)When the system DVD loading as shown in Figure 3-34 on page 3-36 is completed, the system upgrade process is completed successfully.
  • Page 94: Loading The System Software With Usb Memory Stick (For R2.2.2 Or Above)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above) NOTE: Because any data may be lost or impacted during upgrade process, the preserved patient data on the system should be backed up before application software upgrade.
  • Page 95 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above) (cont’d) 6.) The system software load instruction window displays. Select one of the options for loading the system.
  • Page 96 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above) (cont’d) 7.) Input “Yes” or “No” and press Enter key to continue. Figure 3-38 Confirmation on loading the system 3-40 Section 3-9 - Loading the System Software...
  • Page 97 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above) (cont’d) 8.) System USB memory stick will be loaded as shown in Figure 3-39 on page 3-41 .
  • Page 98 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-3 Loading the System Software with USB Memory Stick (For R2.2.2 or above) (cont’d) 9.) After finish updating system, remove the USB memory stick and press any key to shut down the system.
  • Page 99: Functional Check-Out

    ERVICE ANUAL 3-9-4 Functional Check-out 1.) Power on LOGIQ Book XP Series system and wait until system booting to main screen. 2.) Press UTILITY key on control panel 3.) Choose the About button on the right. Figure 3-42 General Screen 4.) Check whether "Software Version"...
  • Page 100: Option Key Update

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-5 Option Key Update NOTE: After the system software loading completion, please input option key and check the option strings to ensure that the options are activated and working. NOTE: Option key is required to be input when the system is powered on after software loading completion.
  • Page 101 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-5 Option Key Update (cont’d) 6.) Some of the options are in disabled status. That means the function of these options is not available. To activate these options, the user need to add new option key. Please select “New“. Figure 3-45 Add new option key 7.) The system pops up a dialogue box for the user to input option key.
  • Page 102 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-5 Option Key Update (cont’d) 8.) The option key is displayed in the SW Option Key list. The status of the options become Valid. That means the function of these options is availabe.
  • Page 103 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-5 Option Key Update (cont’d) The status of activated Options is described in two methods: • Valid until: (date). This status means the option is working only until the displayed date. After the date, the function of this option becomes disabled.
  • Page 104: Probe Recognition Check

    Plug in the probe. In scanning mode, the probe information is displayed on the right top location of the screen. For the probe specification for intended use on LOGIQ Book XP Series, please refer to 3-6-6 on page 3-24.
  • Page 105: Reinstall Dicom Devices

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 3-9-8 Reinstall DICOM Devices Reinstall any DICOM devices used by the customers and check to ensure these DICOM devices work properly. The instruction about installing DICOM devices is not incorporated in this manual. To access the instruction about installing DICOM devices please refer to another manual Basic User Manual.
  • Page 106: Paperwork

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 3-10 Paperwork NOTE: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user information is available during the operation and service of the complete system.
  • Page 107: Functional Checks

    Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ Book XP Series console, diagnostics by using the built-in service software, and power supply adjustments. Table 4-1 Contents in chapter 4...
  • Page 108: General Procedure

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
  • Page 109: Turn On The System

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-1-2 Turn on the system Press the Power On/Off switch at the front of the system once. Power On/Off switch Figure 4-2 Power On/Off Switch When the Power On/Off switch on the Control Panel is pressed once, the Back-end Processor starts and the software code is distributed to initiate the system.
  • Page 110: System Shutdown

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-2-2 System Shutdown Disconnect the Mains Power Cable is necessary. For example: Relocating the system. DO NOT unplug and/or transport the unit until after the power off sequence has been completed. CAUTION Failure to do so may result in corrupted patient files.
  • Page 111: Archiving And Loading Presets

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-3 Archiving and Loading Presets NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible. All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD- R disk for reloading on the system.
  • Page 112: Adjusting The Display Monitor

    6.) Control all stored and residual energy. 7.) Verify isolation. All potentially hazardous stored or residual energy is relieved. Energy Control and Power Lockout for LOGIQ Book XP Series NOTICE WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE GREATER THAN 30 VOLTS: 1.
  • Page 113: System Features

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-6 System Features 4-3-6-1 Control Panel Figure 4-6 Control Panel Tour 1.) TGC 2.) New Patient 3.) End Exam 4.) Mode/Gain/Auto Keys 5.) Preset Key 6.) Imaging/Measurement Keys 7.) Depth 8.) Reverse...
  • Page 114: Logiq Book Xp Series Softmenu Key Tour

    ERVICE ANUAL 4-3-6-2 LOGIQ Book XP Series SoftMenu Key Tour In general, there are three types of softMenu keys: Paddle Switch, Push Button and two-button key. Figure 4-7 SoftMenu Key Tour 1.) The Paddle Switch is used to access the Sub SoftMenu.
  • Page 115: Monitor Display

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-6-3 Monitor Display Figure 4-8 Monitor Display Tour Table 4-2 Monitor Display Features 1. Institution/Hospital Name, Date, Time, Operator 13. Imaging Parameters by Mode (current mode highlighted). Identification, system status (real-time of frozen).
  • Page 116: B Mode Checks

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-7 B Mode Checks 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the system (if it isn’t turned on already) SoftMenu Key Primary Menu...
  • Page 117: B Mode Op Panel Controls

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-7-2 B Mode OP Panel Controls Table 4-3 B Mode Control Panel Controls Step Task Expected Result(s) Press B Mode key B Mode Starts Adjust the field of view. Increasing the depth may view larger/deeper structures rates, and Adjust Depth decreasing the depth may view near the skin line.Press Up/Down Button to increase/ decrease.
  • Page 118: B Mode Softmenu Key

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-7-3 B Mode Softmenu Key Table 4-4 B Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain Adjust Rejection minimum amplitude before it will be processed).
  • Page 119: M Mode Controls

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-8 M Mode Controls 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the system (if it isn’t turned on already).
  • Page 120: M Mode Op Panel Controls

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-8-2 M Mode OP Panel Controls Table 4-5 M Mode OP Panel Controls Step Task Expected Result(s) Press M Mode key M Mode Starts Controls the amount of echo information displayed in an image. Turn B Mode dial to Adjust Gain the left/right to increase/decrease Gain.
  • Page 121: M Mode Softmenu Key

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-8-3 M Mode Softmenu Key Table 4-6 M Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a certain Adjust Rejection minimum amplitude before it will be processed).
  • Page 122: Color Flow Mode Checks

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-9 Color Flow Mode Checks 4-3-9-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector.
  • Page 123: Color Flow Mode Op Panel Controls

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-9-2 Color Flow Mode OP Panel Controls Table 4-7 Color Flow Mode OP Panel Controls Step Task Expected Result(s) Press CFM-Mode key CFM Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
  • Page 124: Color Flow Mode Softemenu Key

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-9-3 Color Flow Mode Softemenu Key Table 4-8 Color Flow Mode Softmenu Key Step Task Expected Result(s) Threshold Threshold assigns the gray scale level at which color information stops. Packet Size Controls the number of samples gathered for a single color flow vector.
  • Page 125: Doppler Mode Checks

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-10 Doppler Mode Checks 4-3-10-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-24, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the system (if it isn’t turned on already).
  • Page 126: Doppler Mode Op Panel Controls

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-10-2 Doppler Mode OP Panel Controls Table 4-9 Doppler Mode OP Panel Controls Step Task Expected Result(s) Press PW Mode key PW Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window. Adjust Gain Turn the Gain dial (PW Mode key) to the left/right to increase/decrease Gain.
  • Page 127: Doppler Mode Op Panel Controls

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-10-3 Doppler Mode OP Panel Controls Table 4-10 Doppler Mode Softmenu Key Step Task Expected Result(s) Selects a level below which echoes will not be amplified (an echo must have a Adjust Rejection certain minimum amplitude before it will be processed).
  • Page 128: Basic Measurements

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-11 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-11-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once; an active caliper displays. 2.) To position the active caliper at the start point (distance) or the most anterior point (tissue depth), move the TRACKBALL.
  • Page 129: Worksheets

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-11-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. The worksheets are selected from the Measurement Touch Panel.
  • Page 130: Using Cine

    Adjust the Frame by Frame up/down Two-Button Softkey to move through CINE memory one frame at a time. 4-3-14 Image Management (QG) For Image Management functionality refer to the LOGIQ Book XP Series Quick Guide. It talks about several topics: • Clipboard •...
  • Page 131: Backup And Restore Database, Preset Configurations And Images

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-15 Backup and Restore Database, Preset Configurations and Images 4-3-15-1 Formatting Media 1.) To format the backup media, CD-RW, select the UTILITY button on the Keyboard. 2.) Select CONNECTIVITY, then TOOLS.
  • Page 132: Backup System Presets And Configurations

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-15-2 Backup System Presets and Configurations NOTE: Always backup any preset configurations before a software reload. This ensures that if the presets need to be reloaded, after the software update, they will be the same ones the customer was using prior to service.
  • Page 133: Restore System Presets And Configurations

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-15-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct CD. 1.) Insert the Backup/Restore CD-R into the drive.
  • Page 134: Archiving Images

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-15-4 Archiving Images 1.) Insert the archive media. To format the archive media, CD-RW, select the Utility button on the Keyboard. 2.) Select Connectivity, then Tools. 3.) Format the CD-RW.
  • Page 135 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 4-3-15-4 Archiving Images (cont’d) Figure 4-22 Image Archive Move Pop-up Menu 7.) Fill in the From Date and To Date. Specify to Keep days together. 8.) Press RECALCULATE.
  • Page 136: Software Configuration Checks

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 4-4 Software Configuration Checks Table 4-11 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct Check that Location (Hospital Name) is correct Location Name is correct Check Language settings...
  • Page 137: Components And Functions (Theory)

    Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains LOGIQ Book XP Series’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes. Table 5-1 Contents in Chapter 5 Section...
  • Page 138: Block Diagrams And Theory

    ERIES ASIC ERVICE ANUAL Section 5-2 Block Diagrams and Theory 5-2-1 Block Diagram Figure 5-1 LOGIQ Book XP Series System Block Diagram • Pre Amp: Preamplifier • OQX2: Beamforming ASIC • CHACOM: B/M/CFM/DOP mode signal processor ASIC • SH-PCI: PCI bridge located between SH4 and PCI bus •...
  • Page 139: General Information

    Doppler A number of combinations of the above • LOGIQ Book XP Series is a digital beam forming system that can handle up to 128 element linear probes. • Signal flow from the Probe Connector Panel to the Front End, to the Mid Processors and Back End Processor and finally to the LCD and peripherals.
  • Page 140: Power Diagrams

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 5-3 Power Diagrams 5-3-1 Overview The AC Power assy’s main tasks are to isolate and output to the DC/DC unit which is inside the system console.
  • Page 141: Air Flow Distribution

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-3-3 Air Flow Distribution Figure 5-3 Air Flow Inside the System The three air flow passes allow the system to be cooled down as shown in the figure above. •...
  • Page 142: Fans

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-3-4 Fans Figure 5-4 Fans The system contains the eight fans at the following positions for producing an air flow. One fan: Inside the CPU for air flow pass C One fan: On the Rear for air flow pass A One fan: On the Bottom for air flow pass B Section 5-3 - Power Diagrams...
  • Page 143: Common Service Platform

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 5-4 Common Service Platform 5-4-1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. The Common Service Platform will increase service productivity and reduce training and service costs.
  • Page 144: The Usage For Security Cable

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-2-4 The usage for security cable The ultrasound system equipped with Kensington security slot which is compatible with a Kensington security cable, refer to Figure 5-6 on page 8 Figure 5-6 How to prevent unauthorized removal of the ultrasound system? 1.) Wrap the cable around the immovable object, refer to...
  • Page 145: Service Home Page

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-3 Service Home Page The navigation bar at the top of the screen allows the user to select from several tools and utilities. Figure 5-8 Customer Service Home Page Chapter 5 Components and Functions (Theory)
  • Page 146: Error Logs Tab (Not Available On Logiq Book Xp Series Now)

    ANUAL 5-4-4 Error Logs Tab (Not available on LOGIQ Book XP Series now) From the Error Logs Tab the Log Viewer displays four categories with pull-down sub-menus and an Exit selection. The Service Interface allows scanner logs to be viewed by all service users.
  • Page 147: Logs

    Log entry headings include Time Stamp; Error Level; Package; Board; Severity; and two (2) new labels that have yet been named. • DICOM* Log entry headings include Time Stamp; Error Level; Package; and Error Message. * This function is option on LOGIQ Book XP Series. NOTICE Chapter 5 Components and Functions (Theory) 5-11...
  • Page 148: Utilities

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-4-2 Utilities The two sub-menus of the Utilities category are Plot Log, and Plot Page. Plot Log Plot Page Figure 5-10 Utilities Sub-Menus • Plot Log Allows for the color coded plot of all Log contents with the package on the ‘x’...
  • Page 149: Exit

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-4-4 Exit The sub-menu, Exit Log Viewer, returns the user to the Service Desktop home page. Figure 5-12 Exit Log Sub-Menu Chapter 5 Components and Functions (Theory) 5-13...
  • Page 150: Diagnostics

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-5 Diagnostics Detailed Diagnostic information is found in Chapter Figure 5-13 User Diagnostic Page 5-4-5-1 Diagnostics Execution Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level diagnostic selection based on the user’s level and login access permissions.
  • Page 151: Image Quality

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-6 Image Quality The Image Quality page is intended to contain tools for troubleshooting image quality issues. FIELD IS NOT YET POPULATED Figure 5-14 Image Quality Page 5-4-7 Calibration The Calibration page is intended to contain the tools used to calibrate the system.
  • Page 152: Configuration

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-8 Configuration The Configuration page is intended to be used to setup various configuration files on the system. The Service Platform is the access and authorization control for remote access to the configuration subsystem.
  • Page 153 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 5-4-11 The PM page is intended to contain the tools used in periodic maintenance of the system. FIELD IS NOT YET POPULATED Figure 5-18 Planned Maintenance Page Chapter 5 Components and Functions (Theory) 5-17...
  • Page 154 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL This page was intentionally left blank. 5-18 Section 5-4 - Common Service Platform...
  • Page 155: Service Adjustments

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the system. These tests are optional. You may use them to check the system for errors.
  • Page 156: Monitor Adjustments

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 6-2 Monitor Adjustments 6-2-1 Adjustments Procedures To adjust the brightness: 1.) Adjust the LCD monitor’s toggle, located beside the Primary and Secondary Control (on the right side of the LCD Monitor).
  • Page 157: Diagnostics/Troubleshooting

    Some Service Tools may be run at the application level. However most software tests are required. Table 7-1 Contents in Chapter 7 Section Description Page Number Overview Gathering Trouble Data Screen Captures Global Service User Interface (GSUI) 7-17 Common Diagnostics 7-19 LOGIQ Book XP Series Diagnostic Descriptions Chapter 7 Diagnostics/Troubleshooting...
  • Page 158: Gathering Trouble Data

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 7-2 Gathering Trouble Data 7-2-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition to be sent back to the manufacturer for analysis.
  • Page 159: Collect A Trouble Image With Logs

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-2-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs: Keyboard Shadow Log Error Logs Crash Log...
  • Page 160: Screen Captures

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 7-3 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key. The P3 key is the factory default print key to accomplish a secondary screen capture.
  • Page 161: Setting The P3 Key To Screen Capture

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-3-1 Check and Record the P3 Key Function (cont’d) 5.) In the Destinations section, record the service that is displayed. The destinations list displays the following information: Name: user defined during service configuration Type: the type of service Destination Device: the device for which the service was configured...
  • Page 162: Capturing A Screen

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-3-3 Capturing a Screen The following is a generic process to capture any screen from the scanner: 1.) Navigate to and display the image/screen to be captured. 2.) Press P3.
  • Page 163: Reset The P3 Key To Customer's Functionality

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-3-3 Capturing a Screen (cont’d) Figure 7-5 Save Dialog Box 6.) A Save dialog box will be opened. Save the image on the Removable disk or CD-R. 7-3-4 Reset the P3 Key to Customer’s Functionality If the customer had programmed the P3 Key to a function other than screen capture, restore that...
  • Page 164: Global Service User Interface (Gsui)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 7-4 Global Service User Interface (GSUI) 7-4-1 Enter global service user interface 1.) When system is running, there is a start icon in the system status bar, see Figure 7-6 on page 7-8 Figure 7-6 system status bar 2.) Click start Icon, appears the selection menu, choose Service Platform icon in the bar, see...
  • Page 165 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-1 Enter global service user interface (cont’d) 3.) Select GE Service in option of User Level of GEMS Service Home page and input correct password, press OK button.
  • Page 166: Active Diagnostic Function

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-2 Active Diagnostic Function 4.) Choose Diagnostic in Global Service Interface to active diagnostic functions, choose options to activate various functions correspondingly. Figure 7-10 Active Diagnostic Function 7-10 Section 7-4 - Global Service User Interface (GSUI)
  • Page 167: Reset Database

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-3 Reset Database 5.) Choose Utilities in Global Service Interface. Click Scanner Utilities. 6.) Choose “Reset Database” to start Database reset tool. Figure 7-11 7.) Click ”Reset Database” to Database reset tool. Chapter 7 Diagnostics/Troubleshooting 7-11...
  • Page 168 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-3 Reset Database (cont’d) 8.) Click “OK” to start reset. 9.) Restart system after completing Database Reset. Figure 7-12 Database Reset 7-12 Section 7-4 - Global Service User Interface (GSUI)
  • Page 169: Clean Userdefs

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-4 Clean Userdefs 10.)Choose Utilities in Global Service Interface. Click Scanner Utilities. 11.)Choose “Clean Userdefs” to start Userdefs clean tool. Figure 7-13 12.)Click ”Clean Userdefs” to Userdefs clean tool. Chapter 7 Diagnostics/Troubleshooting 7-13...
  • Page 170 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-4 Clean Userdefs (cont’d) 13.)Click “OK” to start clean. 14.)Restart system after completing Userdefs Clean. Figure 7-14 Userdefs Clean 7-14 Section 7-4 - Global Service User Interface (GSUI)
  • Page 171: Update Fpga

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-5 Update FPGA 15.)Choose Utilities in Global Service Interface. 16.)Click “LOGIQ Book XP FPGA Update” to active Fpga update tool. Figure 7-15 17.)Click “Update” to Fpga Update tool. 18.)Restart system after completing Fpga upgrade.
  • Page 172: Progress Indicator

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-4-6-3 Progress Indicator Displays a graphical progress indication to the user. 7-4-6-4 Short Text Message Displays either a starting message or aborting message, as well as the diagnostic completion status. 7-4-6-5 Background Color Initially gray, the Control Frame background color changes upon completion of a diagnostic to indicate...
  • Page 173: Common Diagnostics

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 7-5 Common Diagnostics 7-5-1 Utilities Provides two selections: 7-5-1-1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting. 7-5-1-2 System Shutdown Allows for system shutdown from the diagnostic menu. Select to Restart System or Shutdown System. Also, select to retain Disruptive Mode or Not.
  • Page 174: Pc Diagnostics (Non-Interactive Tests)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-5-2 PC Diagnostics (Non-Interactive Tests) 7-5-2-1 CPU Tests 7-5-2-2 Hard Drive Tests 7-5-2-3 Memory Tests 7-5-2-4 CD Drive Test 7-5-2-5 Video Test 7-5-2-6 USB Test 7-5-3 PC Diagnostics (Interactive Tests) 7-5-3-1...
  • Page 175: Logiq Book Xp Series Diagnostic Descriptions

    LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 7-6 LOGIQ Book XP Series Diagnostic Descriptions 7-6-1 Troubleshooting These programs are provided for performing troubleshooting of the system. Table 7-2 Troubleshooting Menu Descriptions Menu Error Code Overall Test Performs all of the troubleshooting tests listed below.
  • Page 176: Fru Test

    Check USCBus USCBus Test 10011 Checks if the SH4 writes the data properly to USC Register or PMX Register. PCI/F1 Test (EUSC) 10012 PCI/F2 Test (IMGD) 10013 Memory Test 10014 7-20 Section 7-6 - LOGIQ Book XP Series Diagnostic Descriptions...
  • Page 177: Dfp

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-6-2-2 These programs are provided for testing the DFP board. Table 7-4 FRU Test, DFP Diagnostics Menu Descriptions Menu Error Code Overall Test Performs all of the DFP board tests listed below. Assy Revision 20001 Test...
  • Page 178: Dtx

    These programs are provided for testing the DTX board. Table 7-5 FRU Test, DTX Diagnostics Menu Descriptions Menu Error Code Overall Test Performs all of the DTX board tests listed below. Assy Revision 30001 Test 7-22 Section 7-6 - LOGIQ Book XP Series Diagnostic Descriptions...
  • Page 179: Dcnn

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 7-6-2-4 DCNN These programs are provided for testing the DCNN board. The PC sends the command to the SH4 in the MST, which brings DCNN into test then gets test report from DCNN via DUSC Bus and sends it back to the PC.
  • Page 180 50002 Memory Memory test 50003 CDRW CDRW test 50004 Video Video test 50005 USB test 50006 PCI Board PCI Board test 50007 AGP Video Card AGP Video Card test 50008 7-24 Section 7-6 - LOGIQ Book XP Series Diagnostic Descriptions...
  • Page 181: Replacement Procedures

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Table 8-1 Contents in Chapter 8 Section Description...
  • Page 182 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 8-2-2 Handle Assy (FRU No. 312) Purpose: This is a description on how to remove and replace the Handle Assy (FRU No.312). 8-2-2-1 Tools • Plier 8-2-2-2 Needed Manpower...
  • Page 183: Handle Assy (Fru No. 312)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 8-2-2 Handle Assy (FRU No. 312) (cont’d) Figure 8-1 Handle Assy DISASSEMBLY Chapter 8 Replacement Procedures...
  • Page 184: Tools

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 8-3 Trackball Roller Cleaning Purpose: This is a description on how to remove the trackball and clean the roller. 8-3-0-1 Tools • No need. 8-3-0-2 Needed Manpower •...
  • Page 185 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 8-3 Trackball Roller Cleaning (cont’d) Figure 8-2 Trackball Roller Cleaning Chapter 8 Replacement Procedures...
  • Page 186 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 8-4Loading the System Software Please refer to Section 3-9 "Loading the System Software" from page 3-27 page 3-49. Section 8-4 - Loading the System Software...
  • Page 187: Renewal Parts

    ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the LOGIQ Book XP Series. Table 9-1 Contents in Chapter 9 Section Description Page Number Overview...
  • Page 188: Renewal Parts Lists

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-3Renewal Parts Lists 9-3-1 Equipment Models Covered in this Chapter Table 9-2 ACDC Power Pack & ACDC Power Cable list Part Name Part Number Description Quantity ACDC Power Pack unit...
  • Page 189: Operator Console Assy

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-4 Operator Console Assy LCD Assy Keyboard Assy Bottom Assy Figure 9-1OPERATOR CONSOLE ASSY Chapter 9 Renewal Parts...
  • Page 190: Keyboard Assy

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-5Keyboard Assy Table 9-3 Keyboard Assy Item Part Name Part Number Description Quantity Trackball 2326353 Trackball Section 9-6Bottom Assy Table 9-4 Bottom Assy Item Part Name Part Number Description...
  • Page 191: Isolation Cart Components

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-7Isolation Cart Components Figure 9-2 Isolation Cart Components Table 9-5 Isolation Cart Components Item Part Name Part Number Description Quantity Wheels 5183729 Wheels Isolation transformer 110V 5180439-2 Isolation transformer 110V Isolation transformer 220V...
  • Page 192 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-5 Isolation Cart Components Item Part Name Part Number Description Quantity Security lock 5183906 Security lock for Twin Isolation Cart (110V) 5177673-2 Isolation Cart (110V) Isolation Cart (220V) 5177558-2 Isolation Cart (220V)
  • Page 193 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-8Isolation Cart Enhanced Version Components Figure 9-3 Isolation Cart Components Chapter 9 Renewal Parts...
  • Page 194: Isolation Cart Enhanced Version Components

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-6 Isolation Cart Components Item Part Name Part Number Description Quantity Isolation Cart (110V) 5384810 Isolation Cart (110V) Isolation Cart (220V) 5384811 Isolation Cart (220V) Isolation cart (110V) 5423269 Isolation cart BT12 (110V)
  • Page 195: Part Number

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-6 Isolation Cart Components Item Part Name Part Number Description Quantity AC Power Cord Switzerland 5177154 AC Power Cord Switzerland AC Power Cord Israel 5176753 AC Power Cord Israel AC Power Cord Argentina...
  • Page 196: Accessories And Kits

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-9Accessories and Kits Table 9-7 Accessories and Kits Item Part Name Part Number Description Battery Pack 2325369-7 Battery Pack 601A Battery Pack 2325369-8 Battery Pack USB CD-RW CN Ver 5251790 DVD-RW with China power cord...
  • Page 197 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-7 Accessories and Kits Item Part Name Part Number Description 614A Color printer (UP-D25MD) CHN Kit 5133106-2 Option part Color printer (UP-D23MD) USA Kit 5133107 Option part 615A...
  • Page 198 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-7 Accessories and Kits Item Part Name Part Number Description PIT Assy for US 5146885 PIT Assy for US PIT Assy for JP 5146887 PIT Assy for JP ECG Module 5129487...
  • Page 199 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-7 Accessories and Kits Item Part Name Part Number Description LOGIQ BOOK XP Software DVD (Chinese 651B 5456030 LOGIQ BOOK XP R2.2.2 China software DVD Version) 651C LOGIQ BOOK XP Software USB...
  • Page 200 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Table 9-7 Accessories and Kits Item Part Name Part Number Description Peripheral Driver Patch CD Installation Kit 5455722 Peripheral Driver Patch CD Installation kit 675A Peripheral Driver Patch USB Installation Kit 5453562 Peripheral Driver Patch USB Installation Kit...
  • Page 201: Manuals

    Quantity LOGIQ Book XP Series Basic Service Manual 5194296-100 Service Manual System User Manuals LOGIQ Book XP Series Basic User Manual - English 5194291-100 Basic User Manual LOGIQ Book XP Series Basic User Manual - French 5194291-101 Basic User Manual...
  • Page 202: Probes

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 9-11Probes Table 9-9 Probes on LOGIQ Book XP Item Part Name Part Number Description Quantity 3C-RS 2290776 Probe (Center Frequency: 3.8MHz) E8C-RS 2290777 Probe (Center Frequency: 6.5MHz) 8C -RS 2354971 Probe (Center Frequency: 6.5MHz)
  • Page 203: Care & Maintenance

    10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ Book XP Series system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory. Some Customers Quality Assurance Programs may require additional tasks and or inspections at a different frequency than listed in this manual.
  • Page 204: Why Do Maintenance

    The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule assumes that you use your LOGIQ Book XP Series for an average patient load (10-12 per day) and use it as a primary mobile unit which is transported between diagnostic facilities.
  • Page 205 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 10-3 Maintenance Task Schedule (cont’d) Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Weekly Monthly QA Program Notes •* Clean Probes * or before each use •* Clean ECG...
  • Page 206: Tools Required

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance Tool Part Number Comments...
  • Page 207: System Maintenance

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-4 System Checks Step Item...
  • Page 208: Functional Checks (See Also Chapter 4)

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-5 System Functional Checks...
  • Page 209: Peripheral/Option Checks

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-6 GE Approved Peripheral/Hardware Option Functional Checks Step...
  • Page 210: Physical Inspection

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Verify that all system labeling is present and in readable condition. Refer to User Manual,..for Labeling details.
  • Page 211: Optional Diagnostic Checks

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-5-6 Optional Diagnostic Checks Optionally you can access the diagnostic software as described in Chapters 5 or 7. View the error logs and run desired diagnostics. 10-5-6-1 View the Logs 1.) Review the system error log for any problems.
  • Page 212: Battery Performance Maintenance

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-5-8 Battery Performance Maintenance It is recommended to do battery performance maintenance one time per year. Please follow the flow chart below to carry out battery performance maintenance. Battery discharge >45 minutes? Battery charge >...
  • Page 213: Electrical Safety Tests

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are based on and conform to IEC 60601-1 Medical Equipment Safety Standards.
  • Page 214: Gems Leakage Current Limits

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-6-2 GEMS Leakage Current Limits The following limits are summarized for IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed. Table 10-11 Chassis Leakage Current Limits—Accessible Metal Surfaces Country Normal Condition...
  • Page 215: Outlet Test - Wiring Arrangement

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-6-3 Outlet Test - Wiring Arrangement Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
  • Page 216: Chassis Leakage Current Test

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-6-4 Chassis Leakage Current Test 10-6-4-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 217: Data Sheet For Enclosure Source Leakage Current

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-6-4-3 Data Sheet for enclosure Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-11. Record all data on the PM Inspection Certificate.
  • Page 218: Probe Leakage Current Test

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-6-5 Probe Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 219: Data Sheet For Transducer Source Leakage Current

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL 10-6-5-4 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-11. Record all data on the PM Inspection Certificate.
  • Page 220: When There's Too Much Leakage Current

    GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL Section 10-7 When There's Too Much Leakage Current... AC/DC FAILS Check any broken of the AC/DC adapter and its cable. Replace a new one if any portion defective. ENCLOSURE FAILS Check any broken of the enclosure.
  • Page 221 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date: Probe 1: Frequency: Scan Format*: Model Number: Serial Number: Probe 2:...
  • Page 222 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) Type BF Applied Part Leakage Current Limits- Probe enclosure Source Leakage Current - Chassis Leakage Current Limits Peripheral 1 Leakage Current Peripheral 2 Leakage Current...
  • Page 223 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL INDEX display location, 4-9 Abbreviations, 9-1 Archiving Images Move, 4-28 Backup Focal zone Patient Database, 4-25 display location, 4-9 Preset Configurations, 4-25 Functional Checks Basic Measurements Basic Measurements, 4-22 Functional Checks, 4-22 Control Panel, 4-7...
  • Page 224 GE H EALTHCARE 5194296-100, R LOGIQ B XP S IRECTION EVISION ERIES ASIC ERVICE ANUAL INDEX Mechanical Safety, 1-9 Restore Media Patient Database, 4-27 Formatting, 4-25 Preset Configurations, 4-27 Monitor Display Functioanl Checks, 4-9 Monitor display location, 4-9 Safety Considerations, 1-9 Move Screen Captures, 7-4 Archiving Images, 4-28...
  • Page 226 2009-2012, General Electric Company. © GE Healthcare-GEMedical Systems Ultrasound. 9900 Innovation Drive Wawautosa, Wisconsin 53226...