Refer to the Module Frames and Modules Technical Manual for detailed information on how to
change these settings.
Webmin
sub-module
Assets Settings
Licensing
ECG Filter
Configuration
STP/TP/ST Settings
P/PT/PP Settings

7.11.1 Module asset settings

NOTE: This configuration applies only to the PDM.
NOTE: The Device Serial Number field is view only and cannot be changed.
To set the device asset number of a PDM:
1.
Log in to Webmin.
2.
Select Configuration > Modules.
3.
In the Sub-Modules for Modules menu, select Assets Settings.
4.
In the Assets Settings window, enter the user-assigned asset number for the device in
the Change Value to column.
5.
Select Save.
The change will take effect immediately.
7.12 Setting the host asset number
Enter a host asset number and view the host serial number.
NOTE: The Host Serial Number field is view only and cannot be changed.
To set the host asset number:
1.
Log in to Webmin.
2.
Select Configuration > Host Asset Settings.
3.
In the Host Asset Settings window, enter the user-assigned host asset number (up to 32
ASCII characters long) in the Change Value to column.
4.
Select Save.
The host asset changes take effect immediately.
7.13 Changing passwords
You can change the passwords for the biomed and clinical users.
Module
PDM
PDM
PDM
E-PRESTN, E-PRETN,
E-RESTN, E-PSM &
E-PSMP
E-P, E-PT & E-PP
Description
This setting allows you to view the customer
assigned asset number of the PDM.
This setting allows you to manage the PDM
feature licenses.
This setting allows you to temporarily
disable the ECG filter of the PDM.
This setting allows you configure the
STP/TP/ST setting after replacing the STP
board.
This setting allows you to configure the
P/PT/PP setting after replacing the STP
board.
Configuration
87
2081903-001