11.3.2 Ping a TCP/IP network device

Use this Webmin feature to verify connectivity with a network device on the MC Network and IX
Network.
1.
Log in to Webmin.
2.
Select Diagnostics > Ping.
3.
In the Address to Ping field in the Ping Command window, type the IP address of a known
device on the network and select ping.
If you receive a reply, then you are able to connect to the device.
If you do not receive a reply, make sure that the patient monitor is connected to an active
network.
The patient monitor withstands a maximum packet loss of 5 packets per 1 million and
maximum latency of 250 ms without performance degradation.
Description
VSYS voltage is measured from the following locations:·
The VSYS ePort voltage is supplied for a PDM module that is
connected to the PDM docking station. It is measured from
the module frame flex board after the VSYS ePort circuit
breaker. The value is shown as "--" if PDM module is not
connected, or if VSYS OPTION voltage is enabled.
The VSYS RACK voltage is measured from the module
frame flex board before the VSYS ePort circuit breaker.
The VSYS EMOD voltage is measured from the E-module
interface board before the VSYS EMOD circuit breaker. The
VSYS EMOD voltage is shown as "--", if the E-module
interface board is not installed.
The VSYS OPTION voltage is supplied for a PDM module
that is connected to the ePort (DB9) interface in the
advanced interface board. It is measured from the
advanced interface board after the circuit breaker. The
value is shown as "--" if PDM module is not connected, or if
VSYS ePort voltage is enabled.
Power consumption is measured from the DC/DC board:
System power is the total power consumption of the
patient monitor taken from the AC/DC output or Battery.
The battery charging power is included in system power by
PMC CPU calculation.
Module power describes the total power consumption of
the connected acquisition modules.
Temperature is measured from the:
CPU board
DC/DC board
E-Module interface board
Advanced interface board
The EMOD and OPTION temperatures are "0", if the related
board is not installed.
Troubleshooting
145
2081903-001