GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 114

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
3.
Read and record the current leakage indicated on the safety tester.
4.
Configure the safety analyzer as follows (SFC):
5.
Read and record the current leakage indicated on the safety tester.
6.
Configure the safety analyzer as follows (SFC):
7.
Read and record the current leakage indicated on the safety tester.
8.
Configure the safety analyzer as follows (SFC):
9.
Read and record the current leakage indicated on the safety tester.
10. Configure the safety analyzer as follows (SFC):
11. Read and record the current leakage indicated on the safety tester.
12. Configure the safety analyzer as follows (NC):
13. Read and record the current leakage indicated on the safety tester.
14. Power off the device under test.
Acceptance criteria in Normal Condition (NC):
All readings shall be less than or equal to 100 μA
Acceptance criteria in Single Fault Condition (SFC) – earth open or one of the supply conductors
open at a time:
106
2081903-001
Polarity: NORMAL
Neutral: OPEN
Earth (GND): CLOSED
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): OPEN
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): OPEN
Polarity: REVERSED
Neutral: OPEN
Earth (GND): CLOSED
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): CLOSED