12.3.4 Detaching the Top Cover

Disassembly and reassembly
1. Press down the Module Frame release knob
and turn the Module Frame to a 90° angle, so
that the Module Frame slot faces to the front.
2. Remove the T10 screw located behind the
Module Frame. [Torque 0.8 Nm]
3. Detach the Top Cover by releasing the four
snaps with a flat blade screwdriver.
Reassemble in reverse order.
Grease the rubber gasket of Top Cover to
reassemble it smoothly.