Detaching The Ac/Dc Board And Rear Unit Assembly - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650

12.3.10 Detaching the AC/DC board and rear unit assembly

Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
d.
12.3.5. Disassembling the main unit into Rear and Front units
DC/DC cable
AC/DC cable
194
2081903-001
1. Detach the DC/DC cable from the cable
holder.
2. Detach the AC/DC cable and the protective
earth cable from the AC/DC Board.
Use pliers, if needed.
3. Remove the T10 screw that attaches the
housing back cover to the Rear Unit Assembly.
[Torque 0.8 Nm]
4.
Release the four snaps that are located
on both side of the AC/DC board.
5. Detach the housing back cover.
NOTE: The loose cable clamp and spring may fall
when detaching the back cover.