Battery Diagnostics - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

The LED is flashing to indicate Activity, i.e., the patient monitor is either transmitting or
receiving data packets over the network.

11.6 Battery diagnostics

You can check the monitor battery status from the Battery Status menu.
1.
Select Monitor Setup > Battery Status
The Monitor Battery Status menu will open.
Field
Time to empty
(hh:mm)
Charge Level (%)
Slot status
Time to full (hh:mm) hh:mm
Temperature
Status
Description
hh:mm
Estimated operating time before the battery is
empty.
N/A
Battery is not connected to the patient monitor
or battery is being charged.
XX
Battery charge level compared to full capacity
(in percentage).
N/A
Battery is not connected to the patient monitor.
"No battery"
Battery is not connected to the patient monitor.
"No communication" Battery is connected to the patient monitor, but
battery communication failure error condition is
on.
"Failure"
Battery error condition is on.
"Discharging"
Patient monitor is operating on battery power.
"Charging"
Patient monitor is connected to AC mains and
charging the battery.
"Full"
Patient monitor is connected to AC mains and
battery is fully charged.
Estimated time to charge the battery to full
capacity.
N/A
Battery is not connected to the patient monitor
or battery is being discharged.
"OK"
Battery temperature is ok.
"Over temperature"
Battery temperature error due to high battery
temperature.
"N/A"
Battery is not connected to the patient monitor.
Troubleshooting