Patient (Source) Leakage Current Tests - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650

8.2.8 Patient (source) leakage current tests

This procedure measures the leakage current from an applied part connector of the device to
ground.
Perform the test in Normal Condition (NC) and in two different Single Fault Conditions (SFC):
1) earth open and 2) one of the supply conductors open at a time.
Perform test with normal and reverse polarity.
NOTE: Refer to the instructions delivered with the safety analyzer to perform this test.
NOTE: *The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characters according to IEC 60601-1.
NOTE: Perform this test for all the connected parameter modules and patient connectors
specified in
1.
Configure the safety analyzer as follows (NC):
2.
Power on the device under test.
3.
Read and record the current leakage indicated on the safety tester.
4.
Configure the safety analyzer as follows (SFC):
5.
Read and record the current leakage indicated on the safety tester.
6.
Configure the safety analyzer as follows (SFC):
7.
Read and record the current leakage indicated on the safety tester.
8.
Configure the safety analyzer as follows (SFC):
108
2081903-001
Table
1.
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): CLOSED
Polarity: NORMAL
Neutral: OPEN
Earth (GND): CLOSED
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): OPEN