12.3.5 Disassembling the main unit into Rear and Front units

Disassemble first:
a.
12.3.2. Detaching the Interface Board
b.
12.3.3. Detaching the Frame Side Housing Decorations
c.
12.3.4. Detaching the Top Cover
188
2081903-001
1. Set the monitor lying on its back and remove
the two screws (T10) from the bottom.
[Torque 0.8 m]
-
NOTE: Be careful not to break the PDM
docking mechanism. Consider detaching it
according to section 12.2.3.
2. Set the rear side of the Main Unit facing you.
3. Remove the three T10 screws from the left
side. [Torque 0.8 Nm]
4. Remove the two T10 screws from the right
side next to the Module Frame hinges. [Torque
0.8 Nm].
5. Detach the Module Frame back plate and
disconnect the Hinge Flex Board from the
Module Frame connectors,
see section 12.2.5.
NOTE: Handle the Hinge Flex Board with care
when disconnected. Do not twist it or bend it
down.
6. Carefully lead the Hinge Flex Board through
the opening, and turn the Module Frame
against the main unit.