Test Completion - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
NOTE: *The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characters according to IEC 60601-1.
NOTE: **According to IEC-60601, the impedance to protect the circuitry and the person
performing the test, but low enough to accept currents higher than the allowable values of the
leakage current to be measured.
WARNING
SHOCK HAZARD - The following step causes high voltage at the test body.
Do not touch the test body.
NOTE: Perform this test for all the connected parameter modules and patient connectors
specified in
1.
Configure the safety analyzer as follows:
2.
Power on the device under test.
3.
Read and record the current leakage indicated on the safety tester.
4.
Configure the safety analyzer as follows:
5.
Read and record the current leakage indicated on the safety tester.
6.
Power off the device under test.
7.
Repeat this test for all the connected parameter modules and patient connectors
specified in
Acceptance criteria:
All readings shall be less than or equal to 50 μA.

8.2.10 Test completion

1.
Disconnect the safety analyzer from the power outlet.
2.
Disconnect the test equipment from the patient monitor.
3.
Disconnect the patient monitor's power cord from the leakage tester.
4.
Mark this task as complete on the Appendix
8.3 Functional check
The purpose of this functional check is to ensure that the system is properly installed and
configured. Cover all peripheral devices that are connected to the patient monitor by
performing the applicable tests below. Skip the tests that are not applicable for the installed
patient monitor.
8.3.1 Start-up
1.
Turn on the patient monitor and perform a cold start.
NOTE: The patient monitor performs a cold start, when the monitor has been off for more than
15 minutes. You can force the patient monitor to perform a cold start earlier by detaching the
110
2081903-001
Table
1.
Polarity: NORMAL
Neutral: CLOSED
GND: CLOSED
Polarity: REVERSED
Neutral: CLOSED
GND: CLOSED
Table
1.
A. Installation check
form.