Detaching The Interface Board - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

12.3.2 Detaching the Interface Board

Reassemble in reverse order.
Disassembly and reassembly
1. Detach the Interface Board cover plate from
the frame by releasing the snaps with a flat
blade screwdriver.
2. Remove the six T10 screws that hold the
Interface Board to the main unit.
[Torque 0.8 Nm]
NOTE: At first reassemble the screws in the
middle.
3. Pull the Interface Board out of the main unit
with pliers.
NOTE! Pull with the pliers from the plate on the
interface board panel.
185
2081903-001

Hide quick links: