Connecting A Secondary Display - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650

6.3 Connecting a secondary display

The patient monitor supports one secondary, clone display.
WARNING
Do not connect a monochrome display to the monitor. Visual alarm
indicators may not appear properly.
WARNING
The secondary display will not sound the audible alarms.
WARNING
To prevent liquids from entering the display, do not tilt the display more
than +/-15 degrees.
WARNING
Use only manufacturer specified mounts.
NOTE: Refer to the patient monitor's supplemental information manual for a list of compatible
secondary displays.
NOTE: Refer to the display's user manual for more information about the display installation.
NOTE: Ensure that the display is installed to the mounting hardware according to the
installation instructions included with the mounting hardware.
NOTE: Make sure that all cables are securely connected.
NOTE: Refer to section
adjustments.
NOTE: All installations should be compliant with IEC 60601-1-1 and local electrical codes.
NOTE: The patient monitor with a non-medical grade display, which is IEC 60950-rated or
equivalent, meets UL and IEC specifications if a medical grade isolation is used. If a
non-medical grade display is to be used, the configuration must meet the IEC 60601-1-1
standard. Refer to IEC 60601-1-1 for requirements if using non-medical grade displays in the
patient environment.
6.3.1 Connections to D15K and D19KT displays
The following instructions describe two methods to connect a D15K or D19KT display as a
secondary display.
A) Cable length is less than 5 meters
1.
Connect one end of the DVI-D to DVI-D video cable to the DVI-I connector on the bottom
of the display and the other end to the DVI-I connector on the back of the patient monitor.
2.
Connect the Type A plug of the standard USB cable to one of the downstream ports
(Type A USB port) on the back of the patient monitor.
3.
Connect the Type B plug of the standard USB cable to the upstream port (Type B USB port)
on the bottom of the display.
60
2081903-001
7.1
for information about touchscreen calibration and display
DVI-I connector
Type B USB port
Type A USB ports