CARESCAPE Monitor B650
Connect a Unity Network ID connectivity device to the network connector labelled "ID" in
the rear panel of the patient monitor.
NOTE: Connection to the Unity Network ID connectivity device is possible only if the monitor is
equipped with the Advanced Interface Board.
WARNING
Before connecting an interfacing module to the device, verify compatibility.
Verify the connectivity of device interfaces before using the equipment.
Verify the compatibility of software versions before using the equipment.
CAUTION
The use of the wrong interface adapter may cause improper operation of the
supported peripheral device.
WARNING
SINGLE PATIENT USE - All eight serial ports of the Unity Network Interface
Device (ID) must only be used on one patient.

6.9 Connecting USB devices

NOTE: The USB-connection to a D15K or D19KT secondary display reserves one of the USB
connectors in the rear panel of the patient monitor.
Connect the following devices to the USB ports in the rear panel of the patient monitor or at
the bottom of a D15K or D19KT display:
Alphanumeric keyboard
Mouse
Remote control
Barcode reader
NOTE: Always use a GE-supplied barcode reader for the CARESCAPE monitors. The barcode
reader comes pre-configured and is ready to connect. Refer to the included installation
instructions for more details.
WARNING
Use only washable keyboard with at least IPX1 protection against ingress of
water.
6.10 Connecting iCollect and other data acquisition
systems
iCollect and other data acquisition systems can be connected to one of the USB connectors in
the rear panel of the patient monitor.
NOTE: Use the ATEN UC232A USB-to-serial Converter and PC serial interface cable to connect
iCollect and the other data acquisition systems to the patient monitor.