12.2.2 Detaching the Recorder Unit

Disassemble first:
a.
12.2.1. Detaching the module frame front cover
Disassembly and reassembly
1. Open the recorder door and remove the paper
roll if installed.
2. Unscrew the two crosshead screws inside the
recorder. Note that the screws cannot be
removed. Torque [1.2 Nm]
NOTE: Be careful not to damage the Hinge Flex
Board on the right of the recorder.
3. Pull the recorder carefully out of the Module
Frame.
NOTE: If you find it difficult to detach you also can
detach the recorder by removing the Modules
Frame cover unit, see section 12.2.4.
4. Release the snaps in the recorder EMC cover
with a flat blade screwdriver while
simultaneously pulling the bottom plate of the
EMC cover.
5. Pull the EMC cover with the recorder board
out of the Module Frame.
179