CARESCAPE Monitor B650
Contact GE Healthcare for any inquiries regarding the software CD and/or activation
code for the new host software version.
NOTE: Loss of monitoring - Software is activated only when the patient monitor is in a patient
discharged / case reset state. Normal patient monitoring is unavailable until the software
activation is completed. This may take up to 10 minutes.
NOTE: The existing clinical and platform settings of the patient monitor are saved and are not
affected by the activation of the new patient monitor software version. However, any new or
changed clinical and platform settings in the activated patient monitor software version have
their factory default values and may require manual configuration. For more information, refer
to the latest version of the CARESCAPE Modular Monitors Supplemental Information Manual.
Contact GE Healthcare for any inquiries regarding the software CD and/or activation code for
the new host software version.
1.
Log in to Webmin.
2.
Select Configuration > Software Management.
The Software Management window displays.
3.
In the Software List, select the software that you want to activate::
Software
Host Software
PDM Software
PiCCO Software
sGAS Software
4.
Select Next.
The selected Software activation window appears.
5.
According to the software you are activating, proceed to:
96
2081903-001
Purpose of use
Patient monitor software for the host unit including:
CARESCAPE Patient Data Module
E-PiCCO module.
CARESCAPE respiratory modules, E-sCAiOV family.
Activating host software immediately
Activating host software after next case end / discharge
Activating module software
UIC SW for the UI board