4.
In Erase Inactive Software After Activation, select:
No
-
to keep the currently active software version as inactive software after
the new software version is activated successfully. This option lets you
restore the patient monitor to the previous software version later.
Yes
-
to erase the currently active software version permanently after the
new software version is successfully activated.
5.
Select Activate to start the software activation.
6.
Confirm activation by selecting Yes in the Host Software window that opens.
The patient monitor shows a message Software activation after next case end / after
next discharge on the display until the next case end or patient discharge. The patient
monitoring can be continued normally until then.
7.
Software activation will start automatically after the next case end / patient discharge,
unless this process is cancelled before it starts.
The patient monitor informs clinical users about pending software activation by showing the
message Software activation after next case end / after next discharge on the display.
The software activation starts automatically after the clinical user performs a patient
discharge / case end the next time. The patient monitor shows a screen saver that informs
about the ongoing software activation:
Software activation in progress. Do not disconnect any measurement modules or other
peripheral devices, or shut down the monitor until the software activation is complete.
Activation may take up to 10 minutes. The device will automatically restart once the
software activation is complete.
NOTE: When the software is activated first time after a uDOM replacement, the screen saver
does not appear.
The clinical user must wait until the software activation is complete and the patient monitor
restarts automatically. If the patient monitor starts up normally and no error messages appear
on the display, the activation is successful.
Activating module software
Follow this instruction when you want to:
activate new sGAS software to a CARESCAPE Respiratory Module
activate new PDM software to a CARESCAPE Patient Data Module
activate new PiCCO software to a E-PiCCO Module.
Before you start:
Make sure that the patient monitor is in a case reset / patient discharge state.
Make sure that the target parameter module is connected to the patient monitor.
NOTE: Do not disconnect the parameter module, or shut down the patient monitor until the
software activation is completed and the parameter module has restarted.
To activate module software:
1.
In the PDM/sGAS/PiCCO Software window, in Current Software State, verify that the
software you want to activate is listed as Inactive.
2.
Select Activate to activate the inactive software.
3.
Confirm activation by selecting Yes in the Host Software window that opens.
The patient monitor will now start activating the module software. It will show a message
"PDM module removed", "Gas measurements removed", or "CO measurement removed"
on the patient monitor screen. This message will remain until the module software
Configuration
99