Ix Network - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

Refer to the sections
WLAN
patient monitor to the S/5 Wireless Network.

5.3.5 IX Network

The IX Network infrastructure shall be installed according to the CARESCAPE Network
Configuration Guide.
The installation site of the patient monitor shall have a wall jack and a network patch
cable for the IX Network.
Refer to the section
configuration information you need to have available to configure the patient monitor to
the IX Network.
Refer to the following sections for the information you need to have available for:
5.4 Installing the mounting hardware
Ensure that all the applicable/ required mounting hardware is properly installed prior to the
installation of the patient monitor:
Mounting hardware for the patient monitor, either for a stand-alone installation or for an
installation to an anesthesia machine or to a ventilator.
Mounting hardware for the PSM module
Mounting hardware for the PDM module
Mounting hardware for the displays
Mounting hardware for the Unity Network ID connectivity device
NOTE: Refer to the CARESCAPE Modular Monitors Mounting Solutions to identify the compatible
mounting hardware for each system component above.
WARNING
Use only manufacturer specified mounts.
5.5 Unity Network ID connectivity device installation
The Unity Network ID connectivity device shall be properly installed, configured and tested
according to the Unity Network ID Connectivity Device Service Manual prior to connecting it to
the patient monitor.
Make sure that the Unity Network ID connectivity device is configured as follows:
IP address is 192.168.253.x, where x is a number between 2 and 254.
Netmask is 255.255.255.0
The location of the Unity Network ID is set to a value other than the default (XXXX-XXX).
For example, BAY3|UNID3+.
Refer to the Unity Network Interface Device (ID) Service Manual for instructions on checking
and changing the IP address.
7.2.3. Selecting and configuring S/5 Network
to see the configuration information you need to have available to configure the
7.2.2. Selecting and configuring CARESCAPE Network
7.5. Configuring printers
7.6. Configuring MUSE/12SL
7.15.1. Configuring the remote service
for IX printer configuration
for Insite with Exc configuration.
Pre-installation requirements
and
7.2.4. Configuring
to see the
2081903-001
53

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents