Login To Webmin - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
IX Netmask: ______________________
4.
Configure the service PC's IP address and subnet mask to the same network segment
with the patient monitor's IX Network.
NOTE: For more information on how to configure the IP address, refer to the PC's
documentation.
5.
Launch a web browser on the service PC.
6.
In the Address field, type https://[IX IP address]:10000 and press Enter.
NOTE: [IX IP address] is the IX Network IP address of the patient monitor.
The Login to Webmin dialog box displays.
7.
Continue to
Closing Webmin:
1.
Restore the original network settings in the service PC.
2.
Disconnect the service PC from the live IX Network.

4.5 Login to Webmin

1.
In the Login to Webmin dialog box, type the username and password and select Login or
press Enter.
Username: biomed
Password: Change<space>Me
44
2081903-001
4.5. Login to
Webmin.