Installing a PDM module to a mounting solution:
Ensure that the selected PDM mount is properly installed according to the installation
instructions included with the mounting hardware.
1.
Connect the PDM module to installed mounting hardware as instructed in the
accompanying installation instructions.
2.
Connect one end of the ePort cable to the PDM module ePort connector.
3.
Connect the other end of the ePort cable to the ePort connector in the rear panel of the
patient monitor.
NOTE: The PDM module can be installed into a pole mount only if the monitor is equipped with
the Advanced Interface Board that provides the ePort connector.
WARNING
PHYSICAL INJURY — Do not install the PDM above a patient. Make sure the
battery is completely inserted and the battery door is completely closed.
Falling batteries could seriously or fatally injure neonatal or other
vulnerable patients.
WARNING
PHYSICAL INJURY — Do not install the PDM above a patient. Leaks from the
battery cells can occur under extreme conditions. The liquid is caustic to the
eyes and skin. If the liquid comes in contact with eyes or skin, flush with
clean water and seek medical attention.
Installing E-modules to the patient monitor
1.
Connect a module by aligning it with the insertion guide inside the module slot.
2.
Push the module into the frame until it locks.

6.5 Connecting to the mains power

WARNING
Use only AC power cords recommended or manufactured by GE.
WARNING
EXCESSIVE LEAKAGE CURRENT - To avoid summation of leakage currents
when interfacing the device with other equipment, the devices may only be
interconnected with each other or to parts of the system when it has been
determined by qualified biomedical personnel that there is no danger to the
patient, the operator, or the environment as a result. In those instances
where there is any element of doubt concerning the safety of the connected
Hardware installation