GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 231

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

Take a snapshot of the refreshed WLAN Diagnostics screen, for example, by printing
it to a file.
3.
Move the roll cart to the following test point along the walking path and repeat the step 2
at each test point until you have completed the test plan. While moving the roll cart from
one test point to another, at all times, verify that there are no losses in the ECG
waveforms.
NOTE: A momentary, up to 5 seconds waveform loss is normal during roaming. If longer,
or repeating waveform losses occur between test points, make this an extra test point
and report it the same way as observations in step 2.
Summary and Reporting
Include the following documents to the test results:
1.
Print the full disclosure reports from the CIC about the observed waveform loss situations.
2.
Print the snapshots of the WLAN diagnostics screens that you saved into the service PC.
3.
Mark to the printouts the id of the test point.
Review and evaluate the test results together with GE personnel and the hospital IT specialists.
Summarize, if additional testing is needed and/or if the WLAN infrastructure needs to be
changed.
Verification procedure for wireless MC Network infrastructure
C- 5 (6)
2081903-001