WARNING
ELECTRIC SHOCK - Always unplug the grounded cables when not in use.
Leaving them connected could result in an electric shock from the ground
contact in the other end.
1.
Turn the monitor off from the
2.
Disconnect the monitor power cord, first from the wall outlet and then from the monitor.
3.
Remove the battery. Unlock the battery cover by turning the battery cover lock 90°
clockwise and pull the battery out.
4.
Disconnect all external cables connected to the monitor.
5.
Detach all parameter modules from the Module Frame.
6.
Detach the monitor from the mount if installed.
NOTE: If you disconnect the monitor from the AC power source and remove the battery prior to
turning the monitor off, the monitor will start alarming about power loss situation. Silence this
alarm by pressing the
When re-installing the battery, make sure that the charge level indicator LEDs are facing
upwards and push the battery in all the way. Close the battery cover lock by turning it 90°
counter-clockwise.
Disassembly workflow
Use this workflow diagram to find the simplest way to disassemble the required parts of the
monitor. Numbers in the diagram refer to the sections in this chapter.
Follow the arrows from the top down to the required part and disassemble the monitor by
following the steps in between.
12.2.1
12.2.3
12.2.2
12.2.4
12.2.5
12.2.6
Silence alarm
12.3.3
12.3.1
12.3.2
12.3.5
12.3.6
12.3.8
12.3.9
12.3.10
12.3.11
On/standby
button.
button.
12.3.4
12.4.2
12.3.7
Disassembly and reassembly
12.4.7
12.4.1
12.4.3
12.4.4
12.4.6
2081903-001
12.4.5
175

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents