GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 201

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

Display
cable
uDOM
UI cable
Reassemble in reverse order.
-
Insert the WLAN card into the card slot.
-
Place the antenna cables carefully inside the cover.
NOTE: Center align the user interface cable carefully to the connector in the CPU Board and/or User Interface
Board.
Note the alignment pins to set the CPU Board in the right place.
NOTE: After replacing the CPU Board, during the first start-up, the patient monitor will automatically check the
EMBC software version in the replaced CPU Board and update the software if necessary. Wait for 5 minutes to
see if the software update is initiated and do not interrupt the process.
Hinge
Flex
cable
Sync
Loudspeaker
cable
cable
Disassembly and reassembly
1. Detach the WLAN card from the CPU if
installed.
NOTE: Be careful not to damage the antenna
cables.
2. Detach the uDOM.
IMPORTANT: Detach the original uDOM and
attach it to the new CPU Board.
The software, licenses and clinical and service
configurations are stored in the uDOM.
3. Detach the Hinge Flex Board and the Display
Cable from the CPU Board.
NOTE: Be careful not to damage the Hinge Flex
Board.
4. Detach the Sync cable, the Loudspeaker
cable, and the U/I cable from the CPU Board.
5. Remove the nine T10 screws that hold the
CPU Board to the mid-frame.
6. Detach the CPU Board. [Torque 0.8 Nm]
193
2081903-001