6.6 Connecting to the MC Network or the S/5 Network

WARNING
MISSED ALARMS - Do not use with iCentral software with Versions 5.0.3 and
earlier or with Mobile Care Server with Version 5.2 and earlier.
WARNING
INCORRECT CALCULATIONS - Using the System with the Aware Gateway
software version 1.4 or earlier could result in incorrect patient height and
weight information. This could lead to incorrect drug dose calculations,
hemodynamic calculations, or oxygenation calculations. Prior to installing
the System, please contact the GE Healthcare Aware Gateway HL7
Integration Engineering Team or your GE Healthcare service representative
to verify or update your Aware Gateway configuration.
WARNING
EXCESSIVE LEAKAGE CURRENT - Only devices that are specified compliant
with IEC 60950-1 or IEC 60601-1 may be connected to the Ethernet MC or IX
ports.
The patient monitor can be connected either to the wired MC Network or S/5 Network.
Connect the MC Network or the S/5 Network patch cable to the network connector
labelled as "MC" in the rear panel of the patient monitor.
6.7 Connecting to the IX Network
Connect the IX Network patch cable to the network connector labelled as "IX" in the rear
panel of the patient monitor.
NOTE: Connection to the IX Network is possible only if the monitor is equipped with the
Advanced Interface Board.
6.8 Connecting to a Unity Network ID connectivity device
Ensure that the Unity Network ID connectivity device is properly mounted, installed and
configured prior to connecting it to the patient monitor.
CAUTION
INSTALLATION - To avoid accidental ingress of liquids, always mount the Unity
Network Interface Device (ID) in a vertical position with the connectors at the
bottom.
Hardware installation
65
2081903-001