Page 1
GE Healthcare Venue 40 Basic Service Manual Part Number: 5314293-100 Revision: 14...
Page 3
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Important Precautions THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY. • IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES.
Page 4
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
Page 5
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES. • KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST. HOIATUS • ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA (ET) TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
Page 6
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN SKILIÐ...
Page 7
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS IZLASĪŠANAS UN SAPRAŠANAS.
Page 8
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
Page 9
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA. •...
Page 10
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
Page 11
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL (ZH-CN) (KO)
Page 12
Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers.
Page 13
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
Page 14
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Revision History Revision Date Reason for change Sep 17th, 2008 Initial Release Feb 26th, 2009 Add FRU replacement procedures, update repair parts list. May 14th, 2009 Update UIs, diagnostics and FRU description Jun 1st, 2009 Update renewal parts list July 27th, 2009...
Typical Users of the Service Manual ......1 - 2 Venue 40 Models Covered by this Manual ......1 - 2 Purpose of Operator Manual(s) .
Page 16
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL CHAPTER 2 Site Preparation Overview ............2 - 1 Purpose of this chapter 2 .
Page 20
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of this chapter 6 .
Page 21
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
Page 22
Disassembly/Re-assembly of Venue 40 ........
Page 23
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
Page 24
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview ............10 - 1 Periodic Maintenance Inspections .
Field Replaceable Units (FRU). 9.) Chapter 9 - Renewal Parts: Contains a complete list of field replaceable parts for the Venue 40. 10.)Chapter 10 - Care & Maintenance: Provides periodic maintenance procedures for the Venue 40.
Venue 40 Console 5418779 Venue 40 Console with film 1-1-6 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the Venue 40 and also kept near the unit for quick reference. Section 1-1 - Overview...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter. Safety Precaution Messages Various levels of safety precaution messages may be found on the equipment and in the service information.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-8 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION Identification and Rating Plate •...
Page 30
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Table 1-8 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION “Protective Earth” indicates the Inside of AC adapter with docking protective earth (grounding) terminal. station “TUV” Listing and Certification Mark is used to designate conformance to nationally recognized product safety standards.
Page 31
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Table 1-8 Warnings LABEL/SYMBOL PURPOSE/MEANING LOCATION Indicates the product contains hazardous materials in excess of the limits established by Chinese standard SJ/T11363-2006 Requirements for Concentration Limits for Certain Hazardous Substances in Electronic Information Products.
Human Safety Operating personnel must not remove the system covers. Servicing should be performed by authorized personnel only. Only personnel who have participated in a Venue 40 Training are authorized to service the equipment. 1-3-3 Mechanical Safety Ultrasound probes are highly sensitive medical instruments that can easily be damaged...
Do not connect the battery to an electrical power outlet. • In the case of the Venue 40 will not be used for a long time, remove the battery. To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 1-3-7 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
Equipment being returned must be clean and free of blood and other infectious substances. GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/ equipment have been properly decontaminated prior to shipment.
1-6-2 CE Compliance The Venue 40 unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 1-7 Customer Assistance 1-7-1 Contact Information If this equipment does not work as indicated in this service manual or in the User Manual, or if you require additional assistance, please contact the local distributor or appropriate support resource, as listed below.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 1-7-2 System Manufacturer Table 1-10 System Manufacturer Manufacturer Phone Number GE Medical Systems (China) Co., Ltd. TEL: +86 510-85225888 No.19, Changjiang Road, Wuxi National Hi-Tech Dev. Zone, Jiangsu, FAX: +86 510-85226688 P.R.China 214028 1-14...
Purpose of this chapter 2 This chapter provides the information required to plan and prepare for the installation of a Venue 40. Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
ASIC ERVICE ANUAL Section 2-2 General Console Requirements 2-2-1 Console Environmental Requirements Table 2-2 Environmental Requirements for Venue 40 Scanners Operational Storage Transport Temperature 10 - 40 degree C -5 - 50 degree C -5 - 50 degree C 30 - 75%...
Page 41
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 2-2-2-2 Inrush Current Table 2-5 Inrush Current Voltage Inrush Current 100V 75A(max) 240V 120A(max) 2-2-2-3 Site Circuit Breaker It is recommended that the branch circuit breaker for the machine be readily accessible. 2-2-2-4 Site Power Outlets A dedicated AC power outlet must be within reach of the unit without extension cords.
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI. The Venue 40 complies with limits as stated on the EMC label. However there is no guarantee that interface will not occur in a particular installation.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 40° C Storage:-10° to 60° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32). SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF NOTICE -10 TO + 60 degrees C.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 2-3 Facility Needs 2-3-1 Recommended Ultrasound Room Layout 2-3-1-1 Purchaser Responsibilities The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system.
Page 46
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 2-3-3-1 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: 18 IN . LINEN SUPPLY (46 C M ) D E D I C AT E D P O W E R R E C EP TA C LE...
Information must include: • A host name, local port number, AE Title, IP address and Net Mask for the Venue 40. • The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
Page 48
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 2-3-4-4 DICOM Option Pre-installation Requirements (cont’d) Venue 40 IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Chapter 3 System Setup Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
Unpacking the scanner 30 minutes Scanner wo/options Dependent on the configuration that is required The Venue 40 installation and functional checkout will take approximately one hour. Venue 40 consoles with optional equipment may take slightly longer. Section 3-1 - Overview...
Do not operate this unit unless all board covers are securely in place. DANGER OPERATOR MANUAL(S) DANGER The User Manual(s) should be fully read and understood before operating the Venue 40 and kept near the unit for quick reference. ACOUSTIC OUTPUT HAZARD DANGER Although the ultrasound energy transmitted from the Venue 40 probe is within FDA limits, avoid unnecessary exposure.
When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1. 3-2-1 Unpacking Venue 40 1.) Cut the four PLASTIC BANDs. 2.) Cut the adhesive tape and open top covers of paper carton.
3.) Take out the paper pad on the side. Refer to Figure 3-4 on page 3-5 4.) Take out the Accessories Package and Console Package. Refer to Figure 3-4 on page 3-5 Figure 3-4 Unpacking Venue 40 Item Description Accessories Package...
Page 54
ENUE ASIC ERVICE ANUAL 3-2-1 Unpacking Venue 40 (cont’d) 5.) Open the Console Package, remove the pad. Take out stylus and battery package. Take out the console. Take off the protective bag and the PE bag. Refer to Figure 3-5 on page 3-6...
Page 55
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-1 Unpacking Venue 40 (cont’d) Item Description Paper pad Battery Stylus Package Battery Bag Foam Venue 40 Protective bag PE bag Chapter 3 System Setup...
Page 56
EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-1 Unpacking Venue 40 (cont’d) 6.) Open the Accessories Package, Figure 3-6 Opening Accessories box Item Description Cover pad Biopsy kit (option) USB cable with miniB Probe bracket for needle guide (option)
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-2 Unpacking Docking Station 1.) Cut the four PLASTIC BANDs. 2.) Cut the adhesive tape and open top covers of paper carton Figure 3-7 Open top covers of paper carton. Chapter 3 System Setup...
Page 58
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-2 Unpacking Docking Station (cont’d) 3.) Remove top foam, take out the Docking Station Module and probe holders. 4.) Take out the Base support module. Figure 3-8 Unpacking the Docking Station Item Description Probe holder bag...
Page 59
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-2 Unpacking Docking Station (cont’d) 5.) Place the Docking Station Module on track of Base Support Module starting from the left, and slowly move it to the right till it clicks: Figure 3-9 Install the Docking Station Chapter 3 System Setup 3-11...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-3 Unpacking Docking Cart If the package of the Docking Cart is a wooden box, 1.) Cut the tape and take off the tool attached to the box. 2.) Insert the ending head of the tool to the hole of the tongue.
Page 61
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-3 Unpacking Docking Cart (cont’d) 7.) Open the box, remove all the foam, cut the wire ties on the chassis. Take out the Docking Cart. Figure 3-11 Unpacking Docking Cart (wooden box) Item Description Power cord bag...
Page 62
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-2-3 Unpacking Docking Cart (cont’d) If the package of the Docking Cart is paper carton, 1.) Cut the adhesive tape 2.) Open the top cover of paper carton 3.) Rotate the plastic locker counter clockwise and pull it out.
Please carefully unpack the system, and do not dispose the package of Venue 40, so that it can CAUTION be reused for service. Please keep the protective bag of Venue 40 in box, so that it can be used for shipping or CAUTION transportation.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up 3-5-1-1 Mount the system to Docking Station/Docking Cart To Mount the system to Docking Station or Docking Cart: 1.) Place the Docking Station and system on a stable surface.
Page 66
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-5-1-1 Mount the system to Docking Station/Docking Cart (cont’d) 3.) Press the locking lever down to the right position Figure 3-15 Press the locking lever 3-18 Section 3-5 - Completing the Installation...
Page 67
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-5-1-2 System Power on 1.) If the Docking Cart is like the following figure A, connect the power cord between Docking module and Cart. If the Docking Cart is like the following figure B, go to step 2. Figure 3-16 Connect power cord between Docking module and Cart Chapter 3 System Setup 3-19...
Page 68
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-5-1-2 System Power on (cont’d) 2.) Plug the AC output connector into the rear panel of Docking Cart. Plug the AC power cord into a grounded, protective earth outlet. Figure 3-17 Connect AC power cord For Docking Station, plug the AC output connector into the AC power input socket, plug the other end into a grounded, protective earth outlet.
Page 69
There is a stylus for touch panel operation. To attach the stylus to system: 1.) Thread the looped end of the stylus strap through the stylus strap post on side of the Venue 40. Figure 3-19 Connect stylus to Venue 40...
Page 70
2.) Thread the stylus through the loop. Figure 3-20 Thread the loop 3.) Place the stylus on the top of the system or in the hole of probe holder. Figure 3-21 Stylus attached to Venue 40 3-22 Section 3-5 - Completing the Installation...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-5-2 Power Off/ Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 3-5-2-1 Back-end Processor Power Down To power down the system:...
3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable damage. 4.) Plug the probe connector into the probe port on right side of Venue 40 with the label facing the front. Figure 3-23 Connect the probe 5.) Carefully position the probe cord so it is free to move and is not resting on the floor.
Page 73
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-5-3 Transducer Connection (cont’d) To disconnect the probe: 1.) Press the locking lever with blue mark to pop up the connector. Figure 3-24 Pop up the locking lever 2.) Pull the probe and connector straight out of the probe port.
ERVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the Venue 40 console are summarized in Figure 3-26 on page 3-26 Table 3-10 Physical Dimensions of Venue 40 Height Width Depth Unit 11.1 10.8...
ANUAL 3-6-4 Connecting Cables Equipment damage possibility. Be sure to use the following recommended connecting cables to WARNING WARNING connect recording devices and a network with Venue 40 console. Table 3-13 List of Connecting Cables Name Part No. Figure NOTE...
ERVICE ANUAL 3-6-5 Peripherals/Accessories Connector Panel Venue 40 peripherals and accessories can be properly connected using the side connector panel. 3-6-5-1 Docking Station/Docking Cart Connector Panel Located on the two sides of Docking Station/Docking Cart are input and output connectors.
Page 78
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-6-5-2 This section indicates the pin assignment for each connector. 1. Pin Assignment of AC input Table 3-14 Pin Assignments of AC input Pin No. Signal Pin No. Signal +20V +20V...
Page 79
3-6-5-3 Connect peripherals 1.) Insert SD Card to system. Pull the SD Socket cover towards the back of the Venue 40 to make it easy to open. Open the SD Card Socket cover. Figure 3-28 Open SD Card Socket cover...
Page 80
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-6-5-3 Connect peripherals (cont’d) 2.) Connect B/W printer to the system. B/W Printer can be properly connected to the USB port of the Docking Station/Cart. Figure 3-30 Connect B/W printer 3.) Connect USB Memory to the system.
Page 81
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-6-5-3 Connect peripherals (cont’d) 4.) Connect external LCD to the DVI port of Docking Station/Docking Cart. Figure 3-32 Connect external LCD 5.) Connect the Wireless Network Card to the USB port of the Docking Station/Cart. Figure 3-33 Connect the Wireless Network Card Chapter 3 System Setup 3-33...
Page 82
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 3-6-5-3 Connect peripherals (cont’d) 6.) Connect the footswitch to the USB port of the Docking Station/Cart. Figure 3-1. Connect the footswitch 3-34 Section 3-6 - System Configuration...
Section 3-7 Software/Option Configuration Refer to the Venue 40 Basic User Manual, Chapter 4, Customizing Your System for information on configuring items like Hospital, Department, Language, Date and Time For information on configuring Software Options, Refer to the Venue 40 Basic User Manual, Chapter 4, Customizing Your System.
Section 8-3 "Loading Base Image Software" on page 8-7 Section 3-9 Software version check-out 1.) Power on Venue 40 scanner and wait until system booting to main screen. 2.) Click Patient using the stylus, press Utility. 3.) Select About in the left column.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 3-10 Paperwork NOTE: During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user information is available during the operation and service of the complete system.
Page 86
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL This page was intentionally left blank. 3-38 Section 3-10 - Paperwork...
Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the Venue 40 console, diagnostics by using the built-in service software, and power supply adjustments. Section Description Page Number Overview...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. 4-3-1 Power On/Boot Up Press Power On/Off key once, the System starts.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-4 System Features 4-3-4-1 Control Panel Figure 4-2 Control Panel Tour 1. Freeze 2. Save 3. Gain 4. Depth 5. Function keys located at the bottom of the system, vary according to different operation modes. NOTE: There is only one depth key on R1.x.x system.
Page 90
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-4-2 Monitor Display Figure 4-3 Monitor Display Tour Table 4-20 Monitor Display Features 1. Hospital/Institute Name 12. Gain 2. Date and Time 13. Acoustic output readout 3. Function selection icons 14.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-5 B Mode Checks 4-3-5-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-35 to the System. 2.) Turn ON the scanner (if it isn’t turned on already) Figure 4-4 B Mode Screen Picture Example Chapter 4 Functional Checks...
Page 92
ASIC ERVICE ANUAL 4-3-5-2 B Mode Control Panel Controls Table 4-21 B Mode Control Panel Controls (on the right side of the Venue 40) Step Task Expected Result(s) Switch to B mode B Mode Starts Adjust the field of view. Increasing the depth...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-6 Color Flow Mode Checks 4-3-6-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-35 to the System. 2.) Turn ON the scanner (if it isn’t turned on already). Figure 4-5 CFM Mode Screen Picture Example Chapter 4 Functional Checks...
Page 94
ASIC ERVICE ANUAL 4-3-6-2 Color Flow Mode Control Panel Controls Table 4-23 Color Flow Mode Control Panel Controls (on the right side of the Venue 40) Step Task Expected Result(s) Switch to PDI mode PDI Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-7 Power Doppler Imaging (PDI) Mode Checks 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-35 to the System. 2.) Turn ON the scanner (if it isn’t turned on already). Figure 4-6 Power Doppler Mode Screen Picture Example Chapter 4 Functional Checks...
Page 96
ASIC ERVICE ANUAL 4-3-7-2 PDI Control Panel Controls Table 4-25 PDI Mode Control Panel Controls (on the right side of the Venue 40) Step Task Expected Result(s) Switch to CF mode CF Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-8 M Mode Checks (R2.x.x, R3.x.x only) 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-35 to the System. 2.) Turn ON the scanner (if it isn’t turned on already). Figure 4-7 Power Doppler Mode Screen Picture Example Chapter 4 Functional Checks 4-11...
Page 98
ASIC ERVICE ANUAL 4-3-8-2 M-Mode Control Panel Controls Table 4-27 M-Mode Control Panel Controls (on the right side of the Venue 40) Step Task Expected Result(s) Switch to B mode B Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-9 Basic Measurements NOTE: The following instructions assume that you acquired an image and then selected Freeze. 4-3-9-1 Distance Measurements 1.) Use the stylus, select Measure, select Distance. 2.) Click on the screen to place the first caliper.
Page 100
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-9-2 Area (Ellipse) Measurements (R2.x.x, R3.x.x Only) 1.) Use the stylus, select Measure, select Ellipse. 2.) Click on the screen to place the first caliper. 3.) Click to place the second caliper, the results will display in the measurement result window in the measurement result window.
Page 101
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-9-3 M-Mode Measurements (R2.x.x, R3.x.x Only) 1.) For R2.0.x: Use the stylus, select Measure, select to perform Tissue Depth measurement Use the stylus, select Measure, select to perform Heart Rate measurement For R2.1.x and R3.x.x: Use the stylus, select Measure, select General tab, select Depth Use the stylus, select Measure, select General tab, select Heart Rate...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-10 Probe/Connectors Usage 4-3-10-1 Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable. 3.) DO NOT allow the probe head to hang free.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-3-11 Using Cine 4-3-11-1 Activating CINE Press Freeze, to activate CINE. 4-3-11-2 Moving through a CINE Loop Frame By Frame Press < or > to move through CINE memory one frame at a time. Figure 4-11 CINE screen example Chapter 4 Functional Checks 4-17...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 4-4 Software Configuration Checks Table 4-29 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct Check that Location (Hospital Name) is correct Location Name is correct Check Language settings Desired Language is displayed...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 4-5-4 Slave USB Port checks Table 4-33 Slave USB port Checks Step Task to do Expected Result(s) Switch the USB switch to miniB USB port Connect PC to the miniB USB port using a USB cable.
Page 106
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL This page was intentionally left blank. 4-20 Section 4-5 - Peripheral Checks...
ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview This chapter explains Venue 40’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes. Table 5-36 Contents in Chapter 5 Section Description...
40 B IRECTION EVISION ENUE AISC ERVICE ANUAL Section 5-2 Block Diagrams and Theory 5-2-1 Block Diagram of Venue 40 and Docking Station/Cart Figure 5-1 Venue 40 and Docking Station System Block Diagram Section 5-2 - Block Diagrams and Theory...
Figure 5-2 Diagram of the Venue 40 (Power flow) When the Venue 40 is running: • If the Venue 40 is running from the battery, the power will go from the battery to the DC/DC Module on the MST Board. •...
Page 110
The TR32 receives the echo from the probe via 64 channels and sends the MLA (multi-line acquisition) data back the demodulate. • Video signals and USB signals communicate between the Venue 40 and the Docking Station/Cart via the Docking port. Section 5-2 - Block Diagrams and Theory...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE AISC ERVICE ANUAL 5-2-3 Block Diagram of Docking Station Figure 5-4 Docking Station Block Diagram • When the Docking Station/Cart is plugged into the power outlet, AC goes into the AC/DC Module. Power flows from the AC/DC Module to the Docking Function Board, then to the Console through the Docking Port.
Page 112
Figure 5-5 Diagram of the Venue 40 (Signal flow) • The USB, DVI and audio signals communicate withe the Venue 40 through the Docking port. The image on the Venue 40 is displayed in the portrait orientation. When connecting to an external monitor through the DVI port, the image will be rotated to the landscape orientation by video rotation board.
Docking Station and this charging circuit takes power from ACDC Module inside the Docking Station.If the AC source is available and battery is installed in the Venue 40, the battery will be charged if not full. This block will also be worked as a battery charging monitor to avoid over heat and over charging of the battery.
Page 114
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE AISC ERVICE ANUAL This page was intentionally left blank. Section 5-3 - Power Diagrams...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
Adjust volume To adjust the volume: 1.) Select Patient, select Utility, select Miscellaneous tab in Setting. 2.) Select Mute, Low, Med or High for Venue 40 speaker and Docking Station/Cart speaker. Figure 6-7 Adjust volume Section 6-2 - Monitor Adjustments...
Page 117
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 6-2-3 Adjust monitor on Docking Station/Docking Cart To adjust the monitor on Docking Station/Docking Cart. 1.) Tilt the LCD monitor for the optimum viewing angle. The maximum angle is 45. Figure 6-8 Tilt the LCD monitor Chapter 6 Service Adjustments...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 6-2-3 Adjust monitor on Docking Station/Docking Cart (cont’d) To adjust the height of Docking Cart. 1.) Hold the cart handle by both hands, step on the pedal, push or lift to adjust the height. Figure 6-9 Cart height adjustment When the cart handles are used for power cable management, the sudden raising of the cart to CAUTION...
Page 119
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 6-3Stylus Stylus tips can be replaced when it becomes dull. To replace the stylus tips: 1.) Use the grooved stylus tip tool in stylus package, grasp and pull out the stylus tip. 2.) To insert a new stylus tip, insert the flat end into the stylus securely using the grooved stylus tip tool.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 7-2Diagnostics 7-2-1 Diagnostic Tools The diagnostic tools are provided for checking the system which includes the power supply, temperature, fan operation, board functions, keyboard operation, peripherals and so on. Before performing diagnostics, please check the software and hardware version in Utility - About.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 7-3Troubleshooting 7-3-1 Console Troubleshooting Trees 7-3-1-1 System Doesn’t Boot This is an overall diagram showing a recommended sequence for troubleshooting at no-boot situation. System doesn’t boot! Power on switch Is the power on LED lit? Is the LCD backlight...
Page 124
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-1 System Doesn’t Boot (cont’d) System doesn’t boot! (con’d) Is Docking Power LED lit? Is system mounted to Docking Station? Replace Docking Cable. Does system boot? Press the battery indicator, is there any LED lit? Go on to the next page.
Page 125
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-1 System Doesn’t Boot (cont’d) System doesn’t boot! (con’d) Is the AC power cord plugged into wall power outlet? Is the wall power Is system mounted to outlet available? Docking Cart? Replace ACDC Module.
Page 126
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-2 Hard Key Low Sensitivity Hard Key Low Sensitivity Take off gloves Wearing Gloves? Still low sen- sitivity Replace Keyboard Assy Still low sen- sitivity Replace Cable Merge Board Cable Still low sen- sitivity...
Page 127
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-3 Structured Artifact in the image Moving artifacts are seen on the LCD! Do they fade out when the scan area is changed? Change the scan area. Does it fade when the probe is re-connected? Do they disappear when...
Page 128
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-3 Structured Artifact in the image (cont’d) Reverberation is seen on the LCD! Does it fade out when the scan area is changed? Change the scan area. Does it fade when the Acoustic output is changed? Change the Acoustic output.
Page 129
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-3 Structured Artifact in the image (cont’d) Reverberation is seen on the LCD! (cont’d) Check the TR32 board assy, refer to TR32 Diagnostic Tests. Does it pass? Swap the TR32 boards.
Page 130
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-4 B Mode Low Sensitivity B Mode is low sensitivity! Is image sensitivity Is sensitivity low for all the recovered after changing probes? probe? Is the transmit power of the Check the HVPS.
Page 131
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-5 B Mode Low Image Quality ROD SCAN Transducer B Mode is low image quality! Scan Area Screw- driver Are there any irregular patterns detected during rod scan? Does it fade when the probe is re-connected?
Page 132
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-6 Noise in B Mode Noise is shown in B Mode. Does the system have proper grounding? Set the functional grounding. Are there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
Page 133
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-7 Color Flow Low Sensitivity Color Flow is low sensitivity. Is the sensitivity normal for B Mode? Go to B Mode Low Sensitivity section. Is the sensitivity low for all types of probe? Try using a different probe Is the sensitivity low for all...
Page 134
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-8 Noise in Color Flow Noise is shown in Color Flow. Are there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices. Is the System Grounded properly? Provide appropriate grounding for...
Page 135
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-9 Stylus - Impaired Sensitivity The stylus is low sensitive! Try using another stylus. Does it work well? Replace the stylus. Are there any electrical devices that affect the stylus? Make sure the stylus is not in close proximity to EMC emitting devices...
Page 136
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-1-10 LCD Display - Impaired Function Error on the LCD display Does the colors seem Replace LCD Cable. abnormal? Still ab- normal Replace Top Assy. Does the Brightness Adjust the brightness.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-2 Peripheral Troubleshooting Trees 7-3-2-1 Unable Recording by Printer Unable to record by the Printer! Is the connection set in the right way? 1) Check power supply connection to the printer.
Page 138
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-2-2 SD Card Unable to save data to SD Card Is SD Card inserted? Insert SD Card. Is Storage Location correctly configured? Change Storage Location to SD Card in Utility.
Page 139
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 7-3-2-3 USB Memory Stick Unable to save data to USB Memory Stick Is the USB Memory Stick connected to the USB port? Connect the USB Memory Stick. Is system properly mounted to Docking Station/Docking Cart?
Page 140
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL This page was intentionally left blank. 7-20 Section 7-3 - Troubleshooting...
If the Venue 40 needs to be sent for repair, ensure that any patient information is erased from the storage device, or that the storage device is removed from the Venue 40 before shipping. In case that any patient information is still residing on the Venue 40, GE will contact the customer and request for urgent collection of that patient information.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-2-3 Standard tools list for Venue 40 Venue 40 Table 8-2 Standard tools list for Part Name Part No. Screw Description Screwdriver Description screw 5162727 Screw M3x25(NL)
2.) Push the trigger to the other side, use the other hand to take the Docking module off track. Refer to Figure 8-1 on page 8-4 Figure 8-1 Docking Station Desk Support Assy disassembly 8-2-4-5 Mounting procedure 1.) Install the new parts in the reverse order of removal. Section 8-2 - Disassembly/Re-assembly of Venue 40...
3.) Reverse the printer shelf with printer, unscrew 4 screws [M3*8], take out the printer. Refer to Figure 8-3 on page 8-6 Figure 8-3 Docking Cart Printer Shelf replacement 8-2-6-5 Mounting procedure 1.) Install the new parts in the reverse order of removal. Section 8-2 - Disassembly/Re-assembly of Venue 40...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 8-3 Loading Base Image Software NOTE: While it is believed to be unnecessary, It would not hurt to remove all transducers. The touch panel is disabled during the upgrading process, do not use stylus to select controls on the NOTICE screen, always press corresponding hard keys to control the process.
Page 148
Figure 8-5 Utility - Setting Table 8-4 Record settings in Utility - Setting Setting General Miscellaneous Storage Video Length Patients On Image Store Live Scan Venue 40 Docking Location in Seconds Screen 2nd ID Printer Enable Area Save Speaker Speaker Footswitch...
Page 149
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-3 Utility - Image Figure 8-6 Utility - Image Table 8-5 Record settings in Utility - Image Image Thermal Name Image Reverse B-Steer+Default NOTE: B-Steer+Default is only available for software version R3.x.x only. Chapter 8 Replacement Procedures...
Page 150
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-4 Utility - Measure (R3.x.x only) Figure 8-7 Utility - Measure Table 8-6 Record settings in Utility - Measure Measure Obstetrics Measure OB Type EFW Format OB Table Measure Study 8-10 Section 8-3 - Loading Base Image Software...
Page 151
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-5 Utility - Connectivity Figure 8-8 Utility - Connectivity Table 8-7 Record settings in Utility - Connectivity Connectivity Wired Wireless Default Network Data IP Address Subnet Mask Gateway Enable DHCP SSID...
Page 152
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x The touch panel is disabled during the upgrading process, do not use stylus to select controls on the NOTICE screen, always press corresponding hard keys to control the process.
Page 153
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 5.) For software version R1.x.x and R2.0.x, select USA, Euro or JPN for the preset. Figure 8-10 Select preset for software version R1.x.x and R2.0.x NOTE: Japan preset is not available on software version R1.0.X.
Page 154
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) For software version R3.x.x, select Global, Euro or JPN for the preset.. Figure 8-11 Select preset for software version R3.x.x 8-14 Section 8-3 - Loading Base Image Software...
Page 155
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 6.) Press Confirm to confirm the preset. Press Exit to cancel. Figure 8-12 Confirm preset for software version R1.x.x and R2.0.x Chapter 8 Replacement Procedures 8-15...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) Figure 8-13 Confirm preset for software version R3.x.x 8-16 Section 8-3 - Loading Base Image Software...
Page 157
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 7.) Press Start to start the upgrading process. Press Exit to cancel this process. Figure 8-14 Startup screen for upgrading process Chapter 8 Replacement Procedures 8-17...
Page 158
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 8.) Before the process is started, the system will count down 10 seconds, the following figure displays the upgrading process.
Page 159
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 9.) If need to pause the upgrade process, press About. 10.)When the process is paused, press Continue to go back to the upgrading, press Exit to cancel the upgrading.
Page 160
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 11.)After the process is completed, press Shutdown to shut down the system. Take out the SD card and reboot the system.
Page 161
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-6 Upgrade process for software R1.x.x, R2.0.x and R3.x.x (cont’d) 12.)Select Patient with stylus, select Utility, select About in the left column, check whether the software version is the right version for use. Figure 8-18 Upgrading process complete 13.)Perform Touch Screen Calibration: go to Patient - Utility - Diagnostic - Miscellaneous, execute touch screen calibration.
Page 162
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-7 Upgrade process for software R2.1.x The touch panel is disabled during the upgrading process, do not use stylus to select controls on the NOTICE screen, always press corresponding hard keys to control the process. Please make sure the battery is installed and fully charged.
Page 163
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-7 Upgrade process for software R2.1.x (cont’d) 5.) Press Start to start the upgrading process. Press Exit to cancel this process. Figure 8-20 Startup screen for upgrading process Chapter 8 Replacement Procedures 8-23...
Page 164
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-7 Upgrade process for software R2.1.x (cont’d) 6.) Before the process is started, the system will count down 10 seconds, the following figure displays the upgrading process. It is not recommended to press Abort during the upgrading process, system crash may occur.
Page 165
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-7 Upgrade process for software R2.1.x (cont’d) 7.) If need to pause the upgrade process, press About. 8.) When the process is paused, press Continue to go back to the upgrading, press Exit to cancel the upgrading.
Page 166
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-7 Upgrade process for software R2.1.x (cont’d) 9.) After the process is completed, press Shutdown to shut down the system. Take out the SD card and reboot the system. If take out the SD card before the system shutdown, system crash may occur.
Page 167
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 8-3-1-7 Upgrade process for software R2.1.x (cont’d) 10.)Select Patient with stylus, select Utility, select About in the left column, check whether the software version is the right version for use. Figure 8-24 Upgrading process complete 11.)Perform Touch Screen Calibration: go to Patient - Utility - Diagnostic - Miscellaneous, execute touch screen calibration.
Page 168
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL This page was intentionally left blank. 8-28 Section 8-3 - Loading Base Image Software...
ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts available for the Venue 40. Table 9-1 Contents in Chapter 9 Section Description Page Number Overview List of Abbreviations...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 9-3Renewal Parts Lists 9-3-1 Power cables 9-3-1-1 Power cables of Docking Cart Table 9-2 Power Cord of Docking Cart Part Name Part Number Description Quantity ACDC Power Cable 5177146 ACDC Power Cable for USA ACDC Power Cable...
Page 171
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 9-3-1-2 Power cables of Docking Station Table 9-4 Power Cord of Docking Station Part Name Part Number Description Quantity ACDC Power Cable 5314913 ACDC Power Cable for USA ACDC Power Cable 5314914 ACDC Power Cable for Europe...
Figure 9-2 Top Assy Table 9-5 Top Assy Item Part Name Part Number Description Quantity Venue 40 Top Assy with LCD, touch screen, front Top assy 5308023 cover, hard keys Venue 40 Top Assy with LCD, touch screen, front 100A Top assy 5308023-2...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 9-6Middle Cover Assy Figure 9-3 Middle Cover Assy Table 9-6 Keyboard Assy Item Part Name Part Number Description Quantity Middle Cover Assy with power key, SD card socket Middle Cover Assy 5321407 and speaker...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 9-8Docking Station Assy 511A 511B Figure 9-5 Docking Station Assy Table 9-8 Docking Station Assy Item Part Name Part Number Description Quantity Docking Probe 5316922 Docking Probe Holder Holder Docking Probe 404A...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 9-9Docking Cart Assy Figure 9-6 Docking Cart Table 9-9 Docking Cart Assy Item Part Name Part Number Description Quantity Wheels 5321852 2 front wheels and 2 back wheels Plastic Shelf 5321901 Plastic Shelf...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 9-10 Accessories and Kits Table 9-10 Accessories and Kits Item Part Name Part Number Description Quantity Stylus Kit 5323661 Stylus Kit SD Card for software 5315797-2 SD Card for software R1.0.3 602A SD Card for software...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Table 9-10 Accessories and Kits Item Part Name Part Number Description Quantity UP-D897 Digital B/W Printer 5151261 UP-D897 Digital B/W Printer (EU) (EU) Chapter 9 Renewal Parts 9-11...
Page 180
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 9-11Manuals Table 9-11 MANUALS Item Part Name Part Number Description Quantity System User Manuals 5265930-100 Basic User Manual English Basic User Manual English 5265930-101 Basic User Manual French Basic User Manual French 5265930-106 Basic User Manual Spanish...
Page 181
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Table 9-11 MANUALS Item Part Name Part Number Description Quantity 5265930-180 Basic User Manual Ukrainian Basic User Manual Ukrainian 5265930-181 Basic User Manual Indonesian Basic User Manual Indonesian 5419428-100 Basic User Manual English for R2.1.x Basic User Manual English for R2.1.x...
10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your Venue 40 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory. Some Customers Quality Assurance Programs may require additional tasks and or inspections at a different frequency than listed in this manual.
The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule assumes that you use your Venue 40 for an average patient load (10-12 per day) and use it as a primary mobile unit which is transported between diagnostic facilities.
Page 185
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Weekly Monthly QA Program Notes •* Clean Probes * or before each use • Inspect AC Mains Cable Mobile Unit Check Weekly •...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
Page 187
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal 9-46696 Hex Keys...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Part Number Comments Digital Volt Meter (DVM) Leakage Current Ultrasound Kit 2113015 For 120V and 220V Units...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-6 System Checks Step Item Description...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 60 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks Table 10-7 System Functional Checks ÷...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to the User Manual for a list of approved peripherals/options. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks Step...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-5-5 Physical Inspection Table 10-11 Physical Checks Step Item Description Verify that all system labeling is present and in readable condition. Refer to User Manual, ..for Labeling details.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-5-6 Optional Diagnostic Checks Optionally you can access the diagnostic software as described in Chapter 5 or 7. View the error logs and run desired diagnostics. 10-5-6-1 View the Logs 1.) Review the system error log for any problems.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-5-8 Battery Performance Maintenance It is recommended to do battery performance maintenance once every year. Please follow the flow chart below to carry out battery performance maintenance. Battery discharge >45 minutes? Battery charge >...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 10-6 Electrical Safety Tests 10-6-1 Safety Test Overview The electrical safety tests in this section are based on and conform to IEC 60601-1 Medical Equipment Safety Standards. They are intended for the electrical safety evaluation of cord-connected, electrically operated, patient care equipment.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-6-2 GEMS Leakage Current Limits The following limits are summarized for IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed. Table 10-13 Chassis Leakage Current Limits—Accessible Metal Surfaces Country Normal Condition...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-6-3 Outlet Test - Wiring Arrangement Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-6-4 Chassis Leakage Current Test 10-6-4-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
Page 199
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-6-4-3 Data Sheet for enclosure Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-13. Record all data on the PM Inspection Certificate. Table 10-16 Typical Data Sheet for enclosure Source Leakage Current Test 2 Tester Neutral...
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-6-5 Probe Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-6-5-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and...
Page 201
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL 10-6-5-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-13. Record all data on the PM Inspection Certificate. Equipment damage possibility.
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Section 10-7 When There's Too Much Leakage Current... AC/DC FAILS Check any broken of the AC/DC cable. Replace with a new one if any portion is defective. ENCLOSURE FAILS Check for any damage to the enclosure.
Page 203
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date: Probe 1: Frequency: Scan Format*: Model Number: Serial Number: Probe 2: Frequency:...
Page 204
GE H EALTHCARE 5314293-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) Type BF Applied Part Leakage Current Limits- Probe enclosure Source Leakage Current - Chassis Leakage Current Limits Peripheral 1 Leakage Current Peripheral 2 Leakage Current...
Page 205
GE H EALTHCARE 5314291-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL INDEX Inrush Current, 2-2, , 2-3 Abbreviations, 9-1 Basic Measurements Functional Checks, 4-13 Boot Up, 3-17 LOTO, 1-11 CE Compliance, 1-12 Mechanical Safety, 1-8 Contact Information, 1-13 Models Covered, 1-2 Control Panel, 4-3 Monitor Display...
Page 206
GE H EALTHCARE 5314291-100, R 40 B IRECTION EVISION ENUE ASIC ERVICE ANUAL Index...
Need help?
Do you have a question about the Venue 40 and is the answer not in the manual?
Questions and answers