Advertisement

CARESCAPE Monitor B650
2.
Continue to
Closing Webmin
Select
4.2 Local access to Webmin with a service PC connected
to the IX connector
You can access Webmin locally by connecting an Ethernet crossover cable between the
service PC and the IX connector of the patient monitor.
NOTE: This connection method is supported only if the monitor is equipped with the Advanced
Interface Board that provides both the IX and the MC connectors.
NOTE: If you disconnect the patient monitor from a live IX Network when a patient is admitted,
you will temporarily lose the services provided by the IX Network, e.g., access to the IX printers
and MUSE reports.
WARNING
Non-medical equipment does not provide the same level of protection
against electrical shock. Do not touch the patient and any part of
non-medical equipment at the same time. Some examples of non-medical
equipment are laser printers and non-medical computers.
1.
Connect a service PC to the IX connector on the patient monitor using a crossover cable.
2.
In patient monitor, select Monitor Setup > Service. The local browser opens and displays
the Login to Webmin dialog box.
3.
Record the IX IP address of the patient monitor:
IX IP address: ______________________
IX Netmask: ______________________
40
2081903-001
4.5. Login to
Webmin.
to close Webmin and return to the main display.