Detaching The E-Module Interface Board - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650

12.2.5 Detaching the E-module Interface Board

Disassemble first:
a.
12.2.1. Detaching the module frame front cover
b.
12.2.3. Detaching the PDM docking mechanism
c.
12.2.4. Detaching the Module Frame cover unit
Reassemble in reverse order.
182
2081903-001
1. Detach the Module Frame back plate.
Release the snaps with a flat blade
screwdriver and detach the plate.
2. Disconnect the Hinge Flex Board from the
Module Frame connectors.
NOTE: Handle the flex board with care when
disconnected. Do not twist it or bend it down.
3. Remove the six T10 screws mounting the
E-module Interface Board. Torque [0.6 Nm]
NOTE: When reassembling, do not use excessive
force to avoid damaging the plastic threads.
4. Detach the board from the Module Frame
unit. Pay attention to the connector on the left
side of the board.