Installing Parameter Modules - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

CARESCAPE Monitor B650
5.
Secure the USB cables connections to the displays according to the instruction included
in the display package.
Video cable:
Connect one end of the 15-meter DVI-I to DVI-I video cable to the DVI-I connector on the
bottom of the display and the other end to the DVI-I connector on the back of the patient
monitor.

6.4 Installing parameter modules

Refer to the patient monitor's supplemental information manual to verify compatible
parameter modules and to identify the parameter modules with identical measurements.
WARNING
Ensure that the compact airway modules and CARESCAPE respiratory
modules are in vertical position when used. Tilting them may result in
erroneous readings.
6.4.1 Installing a PSM or a PDM module
You can install a PSM or a PDM module either to the patient monitor or to a remote mounting
solution.
Connecting a PSM or a PDM module to the patient monitor
1.
Connect a module by aligning it with the insertion guides on the pivoting module frame.
2.
Push the module into the frame until it clicks and stops.
Installing a PSM module to a pole mount:
Ensure that the Pole Mount for PSM is properly installed to an IV pole according to the
installation guide included with the mounting hardware.
WARNING
PHYSICAL INJURY- Take care when mounting devices to an IV pole. If a
device is mounted too high the IV pole may become unbalanced and tip
over.
1.
Connect the PSM module to the Pole Mount for PSM as instructed in the accompanying
installation guide.
2.
Connect the Module Bus Adapter for PSM to the PSM connector in the patient monitor.
3.
Connect the cable from the Pole Mount for PSM to the Module Bus Adapter for PSM.