S/5 Network Communication Issues - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

11.8.8 S/5 Network communication issues

Check the following things before proceeding with any detailed network troubleshooting:
Check status of the wired and wireless network connection indicators and the WLAN
signal strength indicator, for reference see "3.7".
Check status of network status LEDs. Check network compatibility. Refer to the patient
monitor's supplemental information manual.
Check that each patient monitor has been assigned a unique Virtual ID key, see "7.2.3".
Wired S/5 network
Problem
Bedside monitor doesn't
register to the Monitor Network
- all bedside monitors affected.
Bedside monitor doesn't
register to the Monitor Network
– only one patient monitor or
some of the bedside monitors
affected.
Possible cause
The problem is related to the
iCentral, or to the cabling
between the iCentral and the
hub/switch.
The problem is related to the
bedside monitor, or to the
cabling between the bedside
monitor and the hub/switch.
The problem is related to
network setup: the bedside
monitor(s) is/are missing from
the network setup.
The problem is related to view
setup: the bedside monitor(s) is/
are missing from the view setup.
Troubleshooting
Recommended action
Proceed to the iCentral
troubleshooting.
Check the cabling between
the iCentral and the hub/
switch.
Check the patient monitor(s)'
network configuration.
Check the cabling between
the bedside monitors and the
hub.
Check the iCentral network
setup.
Check the iCentral view setup.