7. Select

8.3.9 MC Network and S/5 Network

Perform the following test only if the patient monitor is connected to a wired MC Network or S/5
Network.
NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
1.
Check that a network symbol
screen.
2.
Select Data & Pages – Other Patients and select a patient from the list.
3.
Select View and verify that a window with parameters from another patient displays on
the left side of the screen.
4.
Select Close View to close the window.
8.3.10 Wireless LAN
Perform the following test only if the patient monitor is connected to a wireless MC Network or
S/5 Network.
NOTE: The wireless network must be properly installed and the patient monitor must be within
the wireless coverage area.
Test Procedure
4. Select Scan from Barcode.
5. Scan a known test barcode obtained from the hospital.
NOTE: The barcode data content must be known and in
compliance with the completed parser configuration.
6. Verify that the data content in the barcode is correctly
populated to the related fields in the Patient and the
Administr. Information tabs.
4. Select Edit Name & MRN and press
the Medical Record Number field.
5. Scan a sample barcode that only contains one piece of
information (e.g., a Serial Number barcode from a
module's device label).
6. Verify that the data is correctly populated into the
Medical Record Number field.
to close the open menu and return to the main display.
is displayed in the upper right corner of the
Installation checkout
Enter
to highlight
113