GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 221

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

Module type
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Single Fault Condition (SFC)
Normal Condition (NC)
Patient (sink) leakage current tests
Electrical Safety Test completion
Functional check
Start-up
Display: Picture quality
Display: Touchscreen control
Device Information
Configuration Information
Keypad and
remote: Hard key
Keypad and
remote: Trim Knob
Mouse
Alphanumeric keyboard
Barcode reader
MC Network and S/5 Network
Wireless LAN
S/N
Polarity: NORMAL
Ground closed (normal)
Ground open (normal)
Ground closed (reverse)
Ground open (reverse)
Polarity: Reserved
Polarity: NORMAL
Polarity: Reserved
Observed result
Observed result
Acceptance criteria
Monitor starts up normally.
Text is readable, images are clear and
brightness is good.
Touchscreen is correctly calibrated
and operates correctly.
Device information is correct.
Monitor is correctly configured.
Selected menu opens and selected
activity starts.
Selected menu opens and selected
activity starts.
Selected menu opens and selected
activity starts.
Keyboard language configuration is
correct.
Data is correctly populated.
Network symbol and o
monitor's patient data is correctly
shown on the screen.
WLAN network symbol and o
monitor's patient data is correctly
shown on the screen.
Installation check form
Acceptance
PASS
N.A.
criteria
≤ 10 µA
≤ 50 µA
≤ 50 µA
≤ 50 µA
≤ 50 µA
≤ 10 µA
≤ 50 µA
≤ 50 µA
PASS
N.A.
ther
ther
A - 3 (4)
2081903-001
FAIL
FAIL