Maintenance And Checkout - GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

10 Maintenance and checkout

This chapter specifies the checkout procedure and the maintenance activities to be performed
to the patient monitor after corrective maintenance and during annual planned maintenance.
WARNING
Only perform maintenance procedures specifically described in the manual.
WARNING
Planned maintenance should be carried out annually. Failure to implement
the recommended maintenance schedule may cause equipment failure and
possible health hazards.
NOTE: The manufacturer does not, in any manner, assume the responsibility for performing the
recommended maintenance schedule, unless an Equipment Maintenance Agreement exists.
The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
NOTE: Refer to Module Frames and Modules Technical Manual for corrective and planned
maintenance checkout procedures of the parameter modules.
NOTE: Refer to the PDM section in the Module Frames and Modules Technical Manual for the
battery maintenance procedure of the PDM battery.
Corrective maintenance
Service personnel shall perform the following checkout procedure steps after any corrective
maintenance, before taking the monitor back into clinical use:
Performed service activity
After detaching, replacing or
upgrading:
-
Recorder Unit FRU / Upgrade
-
E-module Board FRU / Upgrade
-
Module Frame Assembly FRU
After detaching or replacing:
-
PDM Docking Mechanism FRU
-
Module Frame Cover Unit Set FRU
After detaching, replacing or
upgrading:
-
Standard Interface Board FRU
-
Advanced Interface Board FRU /
Upgrade
After replacing or upgrading:
-
Wireless LAN FRU / Upgrade
After replacing:
-
Mains Fuses FRU
-
Battery FRU
Required checkout procedure
Visual
Electrical
inspections
safety test
(section 10.1)
(section 10.2)
Yes
Yes
Yes
No
Yes
Yes
Yes
No
Yes
No
Maintenance and checkout
Functional check
(section 10.3)
-
10.3.1 Start-up
-
10.3.3 PSM / PDM
identification
-
10.3.4 E-module identification
-
10.3.13 Recorder
-
10.3.1 Start-up
-
10.3.3 PSM / PDM
identification
-
10.3.1 Start-up
-
10.3.6 Mouse
-
10.3.7 Alphanumeric
keyboard
-
10.3.9 MC Network and S/5
Network
-
10.3.1 Start-up
-
10.3.10 Wireless LAN
-
10.3.1 Start-up
2081903-001
129

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents