GE HEALTHCARE CARESCAPE Monitor B650 Technical Manual page 165

Hide thumbs Also See for CARESCAPE Monitor B650:
Table of Contents

Advertisement

Message
Possible causes
Service Monitor
Software activation failed.
Activation Failed
Setting activation failed.
Module
voltage low
Application
The operating system informs
error:
that the Webmin local
Webmin
browser was terminated
abnormally.
Replace monitor
The full capacity of the battery
battery
is less than or equal to 50%
compared to the full capacity
of a new battery.
Battery failure
Battery failure.
Condition monitor
Battery needs conditioning.
battery
Battery
Battery temperature error due
temperature high
to faulty battery or battery
management error.
Speaker failure
1. Speaker cable loose.
2. Speaker failure.
4. CPU tone generator or audio
One of the supply voltages
for the acquisition
modules is out of
specification.
An analog output voltage
is out of specification.
amplifier failure.
Possible solutions
Reactivate software. If that does not
help, reinstall the software and then
reactivate it.
Reactivate settings.
Log in to Webmin and select
Diagnostics > Hardware Stats to
diagnose the problem. If there is a
faulty board, replace it.
Select Audio Pause to acknowledge
the message.
Replace the monitor battery with a new
one.
Replace battery with a new one.
Condition the monitor battery. See
section
10.4.5. Conditioning a battery
Replace battery with a new one. If that
does not help, replace the DC/DC
board.
1. Connect the speaker cable.
3. Replace the speaker.
5. Replace the CPU board.
Troubleshooting
157
2081903-001