Technical Data - Weinmann BiLevel-ST VENTImotion WM 24800 Service And Repair Instructions

Home ventilation unit
Table of Contents

Advertisement

10. Technical data

Product class as per EC directive
93/42/EEC
Dimensions WxHxD in cm
Weight
Temperature range
– operation
– storage
Air pressure range
Max. respiration pressure at an air
pressure of 600 hPa
Electrical rating
Power consumption during
– operation
– standby
Classification as per
EN 60601-1
– type of protection against
electric shock
– degree of protection against
electric shock
Electromagnetic
compatibility (EMC) as per
EN 60601-1-2:
– radio interference suppression
– radio interference immunity
Mean sound pressure level/oper-
ation as per EN ISO 17510
at a distance of 1 m from the
device in the patient position
Sound pressure level, alarm
IPAP pressure range
EPAP pressure range
CPAP pressure range
Pressure accuracy
Increment
Minimum stable limit pressure
(PLS
of a fault)
Maximum stable limit pressure
(PLS
event of a fault)
Respiratory frequency
Accuracy
Increment
) (min. pressure in the event
min
) (max. pressure in the
max
VENTImotion/VENTIlogic
IIa
23 x 12.5 x 34
approx. 4.5 kg
+5 °C to +35 °C
–40 °C to +70 °C
750 – 1100 hPa
(corresponds to a height of approx. 2500 metres)
25 hPa
115 – 230 V AC, 50–60 Hz
Tolerance -20 % +10 %
230 V
115 V
0.2 A
0.4 A
0.043 A 0.068 A
Protection class II
Type B
EN 55011
EN 61000-3-2, EN 61000-3-3, EN 61000-4-2 to 6,
EN61000-4-8 EN61000-4-11
approx. 35 dB (A) at 20-35 hPa
approx. 31 dB (A) at 20 hPa
approx. 29 dB (A) at 15 hPa
approx. 27 dB (A) at 12 hPa
approx. 25 dB (A) at 10 hPa
approx. 23 dB (A) at
approx. 62 dB (A)
6 to 35 hPa
4 to 20 hPa
4 to 20 hPa
± 0.6 hPa
0.2 hPa
(1 hPa = 1 mbar ≈ 1 cm H
≥ 0 hPa
≤ 60 hPa
6 to 45 l/min
± 0.5 l/min
1 l/min
VENTImotion/VENTIlogic with
VENTIclick
23 x 12.5 x 45.5
approx. 4.8 kg
230 V
115 V
0.35 A
0.7 A
0.043 A 0.068 A
7 hPa
O)
2
55
Technical data

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bilevel-st ventilogic wm 27000