9. Technical Data
9.1 Safe distances
Recommended safe distances between portable or mobile HF telecommunication devices (e.g. mobile
Rated output of
HF device in W
0.01
0.1
1
10
100
Device class EC 93/42
Temperature range: Operation
and storage
Relative humidity: Operation and
storage
Dimensions WxHxD
Weight
Electromagnetic compatibility
according to EN 60601-1-2
– Radio interference suppression
– Radio interference resistance
phones) and data cables
Safe distance as a function of transmission frequency in m
150 KHz - 80 MHz
0.12
0.37
1.17
3.69
11.67
Contact Weinmann for further technical specifications.
VENTIsupport
80 MHz – 800 MHz
0.12
0.37
1.17
3.69
11.67
IIa
+10 °C to +40 °C
0 – 95% non-condensing
135 x 55 x 30 mm
approx. 100 g
EN 55011
(VDE 0875 T11)
EN 61000-4
Parts 2 to 6
800 MHz – 2.5 GHz
0.24
0.74
2.34
7.38
23.33
17