General Power Meter Functions
Setting channel offsets
This gain or loss is applied to the measured power before any mathematical
functions, display offsets or relative functions are included.
Offsets are entered in dB and the range is –100 dB to +100 dB.
A positive value compensates for a loss and a negative value compensates for a
gain.
Procedure
To enter a channel offset:
1 Press
2 Press
3 Use the
4 Press
Offset setting field
Figure 2-10
5 Press
6 Confirm your choice by pressing
68
to display the Channel Setup screen. Confirm the channel
requiring setup is displayed.
to display the Offsets Setup.
Offsets
and
to check the Offset setting field.
Typical channel offset display
to highlight the Offset value field and press
Offset pop-up. Use the numeric keypad to enter the required value in the
Offset pop-up window.
keys to highlight the Offset setting field.
.
dB
Keysight N1911A/1912A User's Guide
Offset value field
to display the