There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9. These tables do not contain any data when the power meter is
shipped from the factory.
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First select the table you want to edit as follows:
1 Press
,
Tables
,
to display the Sensor Tbls screen.
Sensor Cal Tables
Figure 8-11
"Sensor Tbls" screen
2 Choose the table you want to edit using the
and
keys. Press
Ed it Table
to display the Ed it Cal screen as shown in
Figure
8-12.