Power Meter Configuration Changes; Figure 8-1 Auto-Averaging Settings - Keysight N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

8
Using N8480 Series Power Sensors

Power Meter Configuration Changes

The Keysight P-Series power meter recognizes when a Keysight N8480 Series
power sensor is connected. The N8480 Series power sensors (excluding Option
CFT) sensor calibration data is automatically read by the power meter. In addition,
the auto-averaging settings shown in
N8481/2/5/7A
N8486AQ/AR
–1 dBm
–10 dBm
–20 dBm
–30 dBm
Figure 8-1
These values are valid only for the power meter channel connected to a Keysight
NOTE
N8480 Series power sensor. Averaging settings can also be manually configured.
Refer to
232
N8481/2B
0 dBm
30 dBm
29 dBm
20 dBm
10 dBm
0 dBm
–10 dBm
Auto-averaging settings
"Setting Measurement Averaging"
Figure 8-1
Maximum
N8481/2H
Sensor Power
20 dBm
17 dBm
10 dBm
0 dBm
Minimum
Sensor Power
are automatically configured.
Resolution Setting
1
2
3
1
1
2
2
2
4
2
2
4
2
2
16
2
8
128
4
64
256
on page 78 for more details.
Keysight N1911A/1912A User's Guide
4
8
32
32
256
128
512