Recorder Output

The rear panel Recorder Output connectors (1 and 2) produce a dc voltage that
corresponds to the power level in Watts of the channel depending on the
measurement mode. This dc voltage ranges from 0 to +1 Vdc. The output
impedance is typically 1 kΩ. Channel and display offsets, and duty cycle have no
effect on the Recorder Outputs.
For example, the Recorder Outputs can be used to:
Record swept measurements
Level an output from a source using external leveling
Monitor the output power
To access the Recorder menu press
you to switch the Recorder Output signal either on or off. The
Min Power
maximum and 0 Vdc minimum output voltage of the Recorder Output.

Procedure

Set the recorder output as follows:
Ensure you have selected the channel you want to set up.
NOTE
1 Press
2 Use the
3 Press
Keysight N1911A/1912A User's Guide
softkeys allow you to scale the power levels to represent the 1 Vdc
,
Meas Select
and
keys to highlight the Rec o/p: setting field.
to reveal the Recorder option menu.
General Power Meter Functions
, and enable Rec o/p. This allows
to display the Measurement Setup menus.
2
and
Max Power
131