Electromagnetic Compatibility (EMC) Measurements
The low frequency range of the E9304A make it the ideal choice for making EMC
measurements to CISPR (Comite International Special Perturbations
Radioelectriques) requirements, and electromagnetic interference (EMI) test
applications such as the radiated immunity test (IEC61000-4-3).
DC coupling of the E9304A input allows excellent low frequency coverage.
However, the presence of any dc voltages mixed with the signal has an adverse
effect on the accuracy of the power measurement.
The
CAUTION
(5 Vdc) can damage the sensing diode.
Keysight N1911A/1912A User's Guide
E9304A
sensor is DC coupled. DC voltages in excess of the maximum value
Using E9300 E-Series Power Sensors
5
197