Using N8480 Series Power Sensors

N8480 Series Sensors Connection Requirements

Table 8-2
Sensor
N8481A
N8481H
N8482A
N8482H
N8481B
N8482B
N8485A
N8486AR
N8486AQ
N8487A
N8488A
234
N8480 Series connection requirements
Connection Requirements
These power sensors connect directly to the POWER REF.
These power sensors are configured with an attenuator. Prior to
calibration this attenuator must be removed. Replace the attenuator
before making measurements.
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type adapter
(08485-60005) to connect to the POWER REF.Remove this adapter before
making measurements.
These waveguide power sensors have two connectors. Use the
N-Type connector to calibrate the power meter.
This sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) to connect to the POWER REF. Remove this adapter
before making measurements.