8

Using N8480 Series Power Sensors

Introduction

The N8480 Series power sensors is a replacement for the 8480 Series power
sensors (excluding the D-model sensors) with build-in Electrically Erasable
Programmable Read-Only Memory
The N8480 Series power sensors is used for measuring the average power
supplied by RF or microwave source or a device-under-test (DUT). The N8480
Series power sensors places a 50 ohm load on the RF or microwave source. The
power meter indicates the power dissipated in this load in W or dBm.
The N8480 sensors (excluding Option CFT) measure power levels from
–35 dBm to +20 dBm (316 nW to 100 mW), at frequencies from 10 MHz to 33 GHz
and have two independent power measurement path (upper and lower range).
Table 8-1
Sensor
N8481/2/5/7/8A and
N8486AQ/AR
excluding Option CFT
N8481/2B excluding
Option CFT
N8481/2H excluding
Option CFT
[a] Recommended for pulse signals measurement with period of more than one second.
Meanwhile, the N8480 sensors with Option CFT only measure power levels from
–35 dBm to +20 dBm (1 µW to 100 mW) in single range.
[1] The calibration factor table stored in the EEPROM is not applicable for N8480 Series sensors with
230
Power range in the Range setting
Range Setting
AUTO (Default)
LOWER
UPPER
AUTO (Default)
LOWER
UPPER
AUTO (Default)
LOWER
UPPER
Option CFT.
(
EEPROM)
Lower Range
–35 dBm to –1 dBm
–35 dBm to –1 dBm
[a]
-
–5 dBm to +29 dBm
–5 dBm to +29 dBm
[a]
-
–15 dBm to +17 dBm
–15 dBm to +17 dBm
[a]
-
Keysight N1911A/1912A User's Guide
[1]
.
Upper Range
–1 dBm to +20 dBm
-
–35 dBm to +20 dBm
+29 dBm to +44 dBm
-
+29 dBm to +44 dBm
+17 dBm to +35 dBm
-
+17 dBm to +35 dBm